Vitros Digoxin Immunoassay Evaluated for Interference by Digoxin-like Immunoreactive Factors

Authors:

Abstract
Serum digoxin concentrations are routinely monitored because of the narrow therapeutic range of digoxin. Endogenous digoxin-like immunoreactive factors (DLIFs) may cause a false analytic increase in the measured digoxin concentration of patients undergoing digoxin treatment (1)(2). DLIF interference is most commonly found in the sera of neonates, and cord blood and neonatal sera frequently produce false positive results for digoxin (3)(4). Here we examine the newly introduced Vitros digoxin immunoassay (DGXN, Johnson & Johnson Clinical Diagnostics) performed on the Vitros 250 analyzer and compare it to a method (OnLine, Roche Diagnostic Systems) substantially free from DLIF interference (5) and to a method (Digoxin II, Abbott Laboratories) where DLIF interference is known to be present (1). These particular methods are not unique with regard to interference by DLIFs, and methods vary considerably in the impact of DLIF interferences (1)(2)(6)(7).