Small field-of-view dedicated cardiac SPECT systems: impact of projection truncation
Authors:
Open Access
- 1 September 2009
- journal article
- research article
- Published by Springer Nature in European Journal of Nuclear Medicine and Molecular Imaging
- Vol. 37 (3) , 528-536
Abstract
Purpose Small field-of-view (FOV) dedicated cardiac SPECT systems suffer from truncated projection data. This results in (1) neglect of liver activity that otherwise could be used to estimate (and subsequently correct) the amount of scatter in the myocardium by model-based scatter correction, and (2) distorted attenuation maps. In this study, we investigated to what extent truncation impacts attenuation correction and model-based scatter correction in the cases of 99mTc, 201Tl, and simultaneous 99mTc/201Tl studies. In addition, we evaluated a simple correction method to mitigate the effects of truncation. Methods Digital thorax phantoms of different sizes were used to simulate the full FOV SPECT projections for 99mTc, 201Tl, and simultaneous 99mTc/201Tl studies. Small FOV projections were obtained by artificially truncating the full FOV projections. Deviations from ideal heart positioning were simulated by axially shifting projections resulting in more severe liver truncation. Effects of truncation on SPECT images were tested for ordered subset (OS) expectation maximization reconstruction with (1) attenuation correction and detector response modelling (OS-AD), and (2) with additional Monte-Carlo-based scatter correction (OS-ADS). To correct truncation-induced artefacts, we axially extended truncated projections on both sides by duplicating pixel values on the projection edge. Results For both 99mTc and 201Tl, differences in the reconstructed myocardium between full FOV and small FOV projections were negligible. In the nine myocardial segments, the maximum deviations of the average pixel values were 1.3% for OS-AD and 3.5% for OS-ADS. For the simultaneous 99mTc/201Tl studies, reconstructed 201Tl SPECT images from full FOV and small FOV projections showed clearly different image profiles due to truncation. The maximum deviation in defected segments was found to be 49% in the worst-case scenario. However, artificially extending projections reduced deviations in defected segments to a few percent. Conclusion Our results indicate that, for single isotope studies, using small FOV systems has little impact on attenuation correction and model-based scatter correction. For simultaneous 99mTc/201Tl studies, artificial projection extension almost fully eliminates the adverse effects of projection truncation.Keywords
This publication has 37 references indexed in Scilit:
- Validation and Evaluation of Model-Based Crosstalk Compensation Method in Simultaneous$^99rm m$Tc Stress and$^201$Tl Rest Myocardial Perfusion SPECTIEEE Transactions on Nuclear Science, 2004
- Efficient fully 3-D iterative SPECT reconstruction with Monte Carlo-based scatter compensationIEEE Transactions on Medical Imaging, 2002
- Comparison of scatter compensation strategies for myocardial perfusion imaging using Tc-99m labeled sestamibiIEEE Transactions on Nuclear Science, 2002
- Acceleration of Monte Carlo SPECT simulation using convolution-based forced detectionIEEE Transactions on Nuclear Science, 2001
- Modeling respiratory mechanics in the MCAT and spline-based MCAT phantomsIEEE Transactions on Nuclear Science, 2001
- Design and experimental validation of a quantitative myocardial /sup 201/Tl SPECT systemIEEE Transactions on Nuclear Science, 1999
- Application of reconstruction-based scatter compensation to thallium-201 SPECT: implementations for reduced reconstructed image noiseIEEE Transactions on Medical Imaging, 1998
- The design and performance of a simultaneous transmission and emission tomography systemIEEE Transactions on Nuclear Science, 1998
- Quantitative analysis of image truncation in focal-beam CTPhysics in Medicine & Biology, 1993
- Nonuniform attenuation correction using simultaneous transmission and emission converging tomographyIEEE Transactions on Nuclear Science, 1992