Precision of bolus thermodilution cardiac output measurements in patients with atrial fibrillation
- 8 February 2005
- journal article
- clinical trial
- Published by Wiley in Acta Anaesthesiologica Scandinavica
- Vol. 49 (3) , 366-372
Abstract
Background: The precision of bolus thermodilution cardiac output measurements in patients with atrial fibrillation (AF) has not previously been determined. A priori we suspected that the precision would be lower in patients with AF than in patients with sinus rhythm (SR). Consequently, we also determined if the precision could be improved by injecting the thermal indicator into the right ventricle instead of the right atrium.Methods: Cardiac output was determined as the average result of four injections of 10 ml of iced saline. Replicate measurements were performed with thermal indicator injections into the right atrium and ventricle. The coefficients of variation and the precisions were calculated.Results: In the 25 patients with AF, mean cardiac output was 3.96 l min−1 (range 2.4–7.4), the coefficient of variation 0.073 (95% CI ± 0.011), and the precision 0.38 l min−1 (95% CI ± 0.14) with injection into the right atrium. In the 25 patients with SR, mean cardiac output was 4.73 l min−1 (range 2.4–7.3), the coefficient of variation 0.047(95% CI ± 0.006), and the precision 0.38 l min−1 (95% CI ± 0.14). In both groups, an agreement analysis demonstrated that the injection of indicator into the right ventricle resulted in a significantly higher cardiac output [AF+0.25 (95% CI ± 0.15) l min−1, SR+0.29 ( ± 0.20) l min−1].Conclusion: The coefficient of variation for cardiac output determinations is 55% higher in patients with AF. Two measurements, separated by time or intervention, must differ by 15% in AF patients and 9% in SR patients before one can be 95% confident that a real change has taken place.Keywords
This publication has 10 references indexed in Scilit:
- Thermodilution cardiac output – are three injections enough?Acta Anaesthesiologica Scandinavica, 2004
- Thermodilution cardiac outputActa Anaesthesiologica Scandinavica, 2002
- Lack of agreement between thermodilution and carbon dioxide‐rebreathing cardiac outputActa Anaesthesiologica Scandinavica, 2001
- Comparing Methods of Clinical Measurement: Reporting Standards for Bland and Altman AnalysisAnesthesia & Analgesia, 2000
- A Meta-Analysis of Studies Using Bias and Precision Statistics to Compare Cardiac Output Measurement TechniquesJournal of Clinical Monitoring and Computing, 1999
- Errors in the measurement of cardiac output by thermodilutionCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1993
- Thermodilution cardiac outputCritical Care Medicine, 1991
- Measurement of cardiac output by the thermodilution method during left thoracotomy in the lateral position in the dogActa Anaesthesiologica Scandinavica, 1990
- Measurement in Medicine: The Analysis of Method Comparison StudiesJournal of the Royal Statistical Society: Series D (The Statistician), 1983
- Thermal-Dilution TechnicsCirculation Research, 1962