Unintended Consequence of High-Dose Vitamin C Therapy for an Oncology Patient: Evaluation of Ascorbic Acid Interference With Three Hospital-Use Glucose Meters
- 7 June 2020
- journal article
- research article
- Published by SAGE Publications in Journal of Diabetes Science and Technology
- Vol. 15 (4) , 897-900
Abstract
The use of high-dose vitamin C in cancer care has offered promising results for some patients. However, the intravenous (IV) doses used for these patients can reach concentrations that interfere with some strip-based glucose meters. We characterized the impact of vitamin C interference, from standard to the very high doses used for some cancer protocols, using three different hospital-use glucose meters. For two of the three devices tested, increasing concentrations of ascorbic acid caused false elevations in the glucose measurements. The third glucose meter did not provide inaccurate results, regardless of the vitamin C concentration present. Rather, above a certain threshold, the device generated error messages and no results could be obtained.Keywords
This publication has 13 references indexed in Scilit:
- High-dose vitamin C and cancerJournal of Oncological Sciences, 2016
- Performance Evaluation of a Glucose Monitoring System for Point-of-Care Testing With the Critically Ill Patient Population—A Multicenter StudyPoint of Care: The Journal of Near-Patient Testing & Technology, 2015
- Retrospective Evaluation of the Accuracy of Roche AccuChek Inform and Nova StatStrip Glucose Meters when Used on Critically Ill PatientsDiabetes Technology & Therapeutics, 2014
- 18F-FDG PET and PET/CT Patient Preparation: A Review of the LiteratureJournal of Nuclear Medicine Technology, 2014
- Clinical Impact of Sample Interference on Intensive Insulin Therapy in Severely Burned PatientsJournal of Burn Care & Research, 2014
- Analytic Evaluation of a New Glucose Meter System in 15 Different Critical Care SettingsJournal of Diabetes Science and Technology, 2013
- The effects of precision, haematocrit, pH and oxygen tension on point-of-care glucose measurement in critically ill patients: a prospective studyAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2012
- Evaluation of the Impact of Hematocrit and Other Interference on the Accuracy of Hospital-Based Glucose MetersDiabetes Technology & Therapeutics, 2008
- Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testingCritical Care Medicine, 2001
- Effects of Drugs on Glucose Measurements With Handheld Glucose Meters and a Portable Glucose AnalyzerAmerican Journal of Clinical Pathology, 2000