Development of an Automatic Ice Fabric Analyser
Authors:
Open Access
- 1 January 1985
- journal article
- research article
- Published by International Glaciological Society in Annals of Glaciology
- Vol. 6, 281-283
Abstract
An automatic ice fabric analyser has been developed which precisely determines both c- and a-axes of the individual grains in a thin polycrystalline ice specimen by the use of X-ray Laue method (eg Cullity 1956). The method of analysis is computer-aided indexing of the zone-axes in X-ray Laue patterns. Crystal orientations can be automatically determined by the analyser with an error of approximately 0.6_. Time for analysing one specimen containing 100 grains is approximately three hours at present.Keywords
This publication has 2 references indexed in Scilit:
- Computer-Aided Spot Indexing for X-Ray Laue PatternsJapanese Journal of Applied Physics, 1981
- Elements of X-Ray DiffractionAmerican Journal of Physics, 1957