Italian multicenter, prospective study to evaluate the negative predictive value of 16- and 64-slice MDCT imaging in patients scheduled for coronary angiography (NIMISCAD-Non Invasive Multicenter Italian Study for Coronary Artery Disease)
Authors:
- 17 December 2008
- journal article
- research article
- Published by Springer Nature in European Radiology
- Vol. 19 (5) , 1114-1123
Abstract
This was a prospective, multicenter study designed to evaluate the utility of MDCT in the diagnosis of coronary artery disease (CAD) in patients scheduled for elective coronary angiography (CA) using different MDCT systems from different manufacturers. Twenty national sites prospectively enrolled 367 patients between July 2004 and June 2006. Computed tomography (CT) was performed using a standardized/optimized scan protocol for each type of MDCT system (≥16 slices) and compared with quantitative CA performed within 2 weeks of MDCT. A total of 284 patients (81%) were studied by 16-slice MDCT systems, while 66 patients (19%) by 64-slice MDCT scanners. The primary analysis was on-site/off-site evaluation of the negative predictive value (NPV) on a per-patient basis. Secondary analyses included on-site evaluation on a per-artery and per-segment basis. On-site evaluation included 327 patients (CAD prevalence 58%). NPV, positive predictive value (PPV), sensitivity, specificity, and diagnostic accuracy (DA) were 0.91 (95% CI 0.85–0.95), 0.91 (95% CI 0.86–0.95), 0.94 (95% CI 0.89–0.97), 0.88 (95% CI 0.81–0.93), and 0.91 (95% CI 0.88–0.94), respectively. Off-site analysis included 295 patients (CAD prevalence 56%). NPV, PPV, sensitivity, specificity, and DA were 0.73 (95% CI 0.65–0.79), 0.93 (95% CI 0.87–0.97), 0.73 (95% CI 0.65–0.79), 0.93 (95% CI 0.87–0.97), and 0.82 (95% CI 0.77–0.86), respectively. The results of this study demonstrate the utility of MDCT in excluding significant CAD even when conducted by centers with varying degrees of expertise and using different MDCT machines.Keywords
This publication has 42 references indexed in Scilit:
- New Technology for Noninvasive Evaluation of Coronary Artery DiseaseCirculation, 2007
- Accuracy of dual-source CT coronary angiography: first experience in a high pre-test probability population without heart rate controlEuropean Radiology, 2006
- Accuracy of 64-MDCT in the Diagnosis of Ischemic Heart DiseaseAmerican Journal of Roentgenology, 2006
- Heart Disease and Stroke Statistics—2006 UpdateCirculation, 2006
- Usefulness of Multidetector Row Spiral Computed Tomography With 64- × 0.6-mm Collimation and 330-ms Rotation for the Noninvasive Detection of Significant Coronary Artery StenosesThe American Journal of Cardiology, 2006
- Diagnostic Accuracy of Noninvasive Coronary Angiography Using 64-Slice Spiral Computed TomographyJACC, 2005
- Limited diagnostic yield of non-invasive coronary angiography by 16-slice multi-detector spiral computed tomography in routine patients referred for evaluation of coronary artery diseaseEuropean Heart Journal, 2005
- Non-invasive coronary angiography with high resolution multidetector-row computed tomography. Results in 102 patientsEuropean Heart Journal, 2002
- Executive Summary of the Third Report of the National Cholesterol Education Program (NCEP) Expert Panel on Detection, Evaluation, and Treatment of High Blood Cholesterol in Adults (Adult Treatment Panel III)JAMA, 2001
- A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart AssociationCirculation, 1975