Measurement of trapping time constants in proton-irradiated silicon pad detectors
Authors:
- 20 December 2004
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 51 (6) , 3055-3062
Abstract
Silicon pad-detectors fabricated from oxygenated silicon were irradiated with 24-GeV/c protons with fluences between 2/spl middot/10/sup 13/ n/sub eq//cm/sup 2/ and 9/spl middot/10/sup 14/ n/sub eq//cm/sup 2/. The transient current technique was used to measure the trapping probability for holes and electrons. The measured trapping probabilities scale linearly with the fluence. Annealing, accelerated at 60/spl deg/C, leads to an increased trapping for holes while electron trapping decreases.Keywords
This publication has 6 references indexed in Scilit:
- Effective trapping time of electrons and holes in different silicon materials irradiated with neutrons, protons and pionsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2002
- Radiation hard silicon detectors—developments by the RD48 (ROSE) collaborationNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2001
- A new method of carrier trapping time measurementNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2000
- The evolution of the MINIMOS mobility modelSolid-State Electronics, 1990
- Intrinsic Optical Absorption in Single-Crystal Germanium and Silicon at 77°K and 300°KPhysical Review B, 1955
- Currents Induced by Electron MotionProceedings of the IRE, 1939