Test‐retest reliability and reproducibility of short‐echo‐time spectroscopic imaging of human brain at 3T
Open Access
- 28 February 2011
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 66 (2) , 324-332
Abstract
A1H magnetic resonance spectroscopic imaging study at 3T and short echo time was conducted to evaluate both the reproducibility, as measured by the interscan coefficient of variation (CV), and test‐retest reliability, as measured by the intraclass correlation coefficient (ICC), of measurements of glutamate (Glu), combined glutamate and glutamine (Glx), myo‐inositol (mI),N‐acetylaspartate, creatine, and choline in 21 healthy subjects. The effect of partial volume correction on these measures and the relationship of reproducibility and reliability to data quality were also examined. A1H magnetic resonance spectroscopic imaging slice was prescribed above the lateral ventricles and single repeat scans were performed within 30 min to minimize physiologic variability. Interscan CVs based on all the voxels varied from 0.05 to 0.07 forN‐acetylaspartate, creatine, and choline to 0.10–0.13 for mI, Glu, and Glx. Findings on the reproducibility of gray and white matter estimates ofN‐acetylaspartate, creatine, and choline are consistent with previous studies using longer echo times, with CVs in the range of 0.02–0.04 and ICC in the range of 0.65–0.90. CVs for Glu, Glx, and mI are much lower than reported in previous studies at 1.5T, while white matter mI (CV = 0.04, ICC = 0.93) and gray matter Glx (CV = 0.04, ICC = 0.68) demonstrated both high reproducibility and test‐retest reliability. Magn Reson Med, 2011.Keywords
Funding Information
- The Mind Research Network (DOE DE-FG02-99ER62764)
- National Institutes of Health (P20 RR021938-01)
This publication has 31 references indexed in Scilit:
- Genetic variation influences glutamate concentrations in brains of patients with multiple sclerosisBrain, 2010
- Reproducibility of serial whole‐brain MR Spectroscopic ImagingNMR in Biomedicine, 2010
- Neurometabolite Concentrations in Gray and White Matter in Mild Traumatic Brain Injury: An 1 H–Magnetic Resonance Spectroscopy StudyJournal of Neurotrauma, 2009
- Quantitative spectroscopic imaging with in situ measurements of tissue water T 1 , T 2 , and densityMagnetic Resonance in Medicine, 2009
- Comparative reliability of proton spectroscopy techniques designed to improve detection of J‐coupled metabolitesMagnetic Resonance in Medicine, 2008
- Reproducibility study of whole‐brain 1H spectroscopic imaging with automated quantificationMagnetic Resonance in Medicine, 2008
- Automatic repositioning of MRSI voxels in longitudinal studies: Impact on reproducibility of metabolite concentration measurementsJournal of Magnetic Resonance Imaging, 2008
- Robust analysis of short echo time 1H MRSI of human brainMagnetic Resonance in Medicine, 2006
- Detection of glutamate in the human brain at 3 T using optimized constant time point resolved spectroscopyMagnetic Resonance in Medicine, 2005
- Estimation of metabolite concentrations from localized in vivo proton NMR spectraMagnetic Resonance in Medicine, 1993