How to Ward Off Retinal Toxicity of Perfluorooctane and Other Perfluorocarbon Liquids?
Authors:
Open Access
- 1 October 2018
- journal article
- research article
- Published by Association for Research in Vision and Ophthalmology (ARVO) in Investigative Ophthalmology & Visual Science
- Vol. 59 (12) , 4841-4846
Abstract
Purpose: Reactive and underfluorinated impurities are acknowledged as a source of cytotoxicity of perfluorocarbon liquids (PFCLs) used as blood substitutes. To determine whether this is also a relevant factor in retinal toxicity, we analyzed eight PFO batches associated with adverse ocular events. Methods: (A) The amount of reactive and underflurinated impurities was analyzed by fluoride-selective potentiometry and expressed as H-value. (B) Cytotoxicity of these batches was determined by an ISO 10993-5-compliant extractive test and compared to published data generated with a direct-contact method. (C) A toxic PFO batch (061014) was purified to remove reactive and underfluorinated impurities. (A) and (B) -measurements were repeated after that. (D) The dose dependence of the H-value and cytotoxicity was determined in a dilution experiment. Results: (A) The batches revealed H-values ranging from 1.400 ppm to 4.500 ppm. (B) All batches induced cell growth inhibition; seven must be classified as cytotoxic. Findings from ISO-conform extractive and direct-contact methods showed no difference. (C) After all reactive and underfluorinated impurities in batch 061014 were removed, the H-value dropped to Conclusions: The clinical relevance of the H-value as a safety parameter for PFO endotamponades could be proven. The H-value is a measure for reactive and underfluorinated impurities that cause toxicity of PFCLs and should be incorporated in each endotamponade specification with a limit of 10 ppm to prove the effectiveness of the ultra-purification required and ensure a safe product. Despite the fact that an (ISO)-standard literally is a “standard” only, which cannot cover all imaginable possibilities, the incorporation of the H-value determination into the relevant ISO standard has been initiated. If a thorough risk assessment results in risks that cannot be detected and/or managed by the effective standard, additional investigations have to be performed.Keywords
This publication has 11 references indexed in Scilit:
- BLINDNESS RELATED TO PRESUMED RETINAL TOXICITY AFTER USING PERFLUOROCARBON LIQUID DURING VITREORETINAL SURGERYRetina, 2018
- Comparison between direct contact and extract exposure methods for PFO cytotoxicity evaluationScientific Reports, 2018
- ACUTE RETINAL DAMAGE AFTER USING A TOXIC PERFLUORO-OCTANE FOR VITREO-RETINAL SURGERYRetina, 2017
- Impure Perfluorocarbon LiquidsRetina, 2017
- Vitreous Substitutes: The Present and the FutureBioMed Research International, 2014
- Efficacy of Short-term Postoperative Perfluoro-n-octane Tamponade for Pediatric Complex Retinal DetachmentAmerican Journal of Ophthalmology, 2014
- Perfluorocarbon Liquid: Its Application in Vitreoretinal Surgery and Related Ocular InflammationBioMed Research International, 2014
- Vitreous Substitutes: A Comprehensive ReviewSurvey of Ophthalmology, 2011
- Vitreo-retinal Surgery2007
- Blutersatzstoffe auf FluorcarbonbasisNachrichten aus Chemie, Technik und Laboratorium, 1990