A Comparative Study on Depth of Penetration Measurements in Diagnostic Ultrasounds Through the Adaptive SNR Threshold Method
Open Access
- 2 March 2023
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 72 (00189456) , 1-8
Abstract
Depth of penetration (DOP) has been investigated in the scientific literature as an informative parameter able to monitor over time both the sensitivity and the general performance of ultrasound (US) diagnostic systems. In common practice, this parameter may suffer from operator-related errors due to its visual assessment. Different image analysis algorithms have been proposed in the literature to address this issue. In this regard, this work evaluates the adaptive SNR threshold method (AdSTM) on six US diagnostic systems equipped with three US probe models, operating at four frequencies. Data were collected from a US phantom with two distinct zones with different attenuation coefficients. The AdSTM results were compared to the outcomes provided by the naked eye method (NEM), which was performed by five non-medical observers. Despite the small population sample of observers, the obtained results were generally consistent across methods, and suggest the implementation of a calibration procedure for AdSTM, and more extensive testing.Keywords
This publication has 22 references indexed in Scilit:
- A Comparative Study on the Influence of Probe Placement on Quality Assurance Measurements in B-mode Ultrasound by Means of Ultrasound PhantomsThe Open Biomedical Engineering Journal, 2015
- A novel approach to a phantom based method for maximum depth of penetration measurement in diagnostic ultrasound: a preliminary study2015
- Guideline for Technical Quality Assurance (TQA) of Ultrasound devices (B-Mode) – Version 1.0 (July 2012)Ultraschall in der Medizin - European Journal of Ultrasound, 2012
- Criteria and scheduling of quality control of b‐mode and doppler ultrasonography equipmentJournal of Clinical Ultrasound, 2012
- Quality assurance in diagnostic ultrasoundEuropean Journal of Radiology, 2011
- Four-Year Experience with a Clinical Ultrasound Quality Control ProgramUltrasound in Medicine & Biology, 2011
- Objective Performance Testing and Quality Assurance of Medical Ultrasound EquipmentUltrasound in Medicine & Biology, 2007
- Increasing the Number of Gray Shades in Medical Display Systems—How Much is Enough?Journal of Digital Imaging, 2006
- Implementation and validation of three automated methods for measuring ultrasound maximum depth of penetration: Application to ultrasound quality controlMedical Physics, 2005
- Real‐time B‐mode ultrasound quality control test procedures. Report of AAPM Ultrasound Task Group No. 1Medical Physics, 1998