Abstract
We report results of a retrospective review of intra-aortic balloon pump (IABP) use in two Australasian centres and evaluate the effect of final IABP tip position on outcome. Indications for counterpulsation, patient demographics and in-hospital outcomes and complications were retrospectively collected. The chest X-ray reports provided the ‘final’ position of the IABP tip. The position was defined as acceptable (tip was seen just below the aortic arch, at T2–T5 vertebrae), malpositioned (tip >5 cm below aortic arch or at T5–T6) or severely malpositioned (tip >10 cm below aortic arch or at T7 or below). Major complications were considered a composite of death secondary to IABP, major limb ischaemia, major IABP malfunction, balloon rupture or haemorrhage, severe renal dysfunction (rise in creatinine >200 μmol/l), stroke and mesenteric ischaemia. Six hundred and forty-five cases were reviewed. The overall major complication rate was 26.2% and 24.3%. Severe renal impairment was the most common complication (16.6%), and second, severe catheter dysfunction (5.4%). Final IABP position was acceptable in 39.9%, malpositioned in 11.1%, severely malpositioned in 6.7% and unavailable for 42.4%. Logistic regression analysis showed IABP tip malposition (compared with satisfactory position odds ratio=3.9 [95% confidence interval=2.0–7.6, P P P <0.001). The acceptance of a less than ideal final position was highly predictive of morbidity directly related to IABP device therapy.Keywords
This publication has 19 references indexed in Scilit:
- Intra-aortic balloon counterpulsation in acute myocardial infarction complicated by cardiogenic shock (IABP-SHOCK II): final 12 month results of a randomised, open-label trialThe Lancet, 2013
- Percutaneous left ventricular assist devices vs. intra-aortic balloon pump counterpulsation for treatment of cardiogenic shock: a meta-analysis of controlled trialsEuropean Heart Journal, 2009
- Role of Prophylactic Intra-Aortic Balloon Pump in High-Risk Patients Undergoing Percutaneous Coronary InterventionThe American Journal of Cardiology, 2006
- A randomized multicenter clinical study to evaluate the safety and efficacy of the TandemHeart percutaneous ventricular assist device versus conventional therapy with intraaortic balloon pumping for treatment of cardiogenic shockAmerican Heart Journal, 2006
- Influence of thrombolytic therapy, with or without intra-aortic balloon counterpulsation, on 12-month survival in the SHOCK trialAmerican Heart Journal, 2003
- Elective versus provisional intra-aortic balloon pumping in high-risk percutaneous transluminal coronary angioplastyAmerican Heart Journal, 2003
- Stroke Complicating Percutaneous Coronary InterventionsCirculation, 2002
- Lack of progress in cardiogenic shock: lessons from the GUSTO trialsEuropean Heart Journal, 2000
- Influence of Catheter and Arterial Diameter on Flow Distal to an Intra-Aortic Balloon Insertion SiteASAIO Journal, 1998
- Identification of patients at high risk for complications of intraaortic balloon counterpulsation: A multivariate risk factor analysisThe American Journal of Cardiology, 1984