Dynamic light scattering biosensing based on analyte-induced inhibition of nanoparticle aggregation
- 12 April 2020
- journal article
- research article
- Published by Springer Nature in Analytical and Bioanalytical Chemistry
- Vol. 412 (14) , 3423-3431
Abstract
A new approach to direct quantitative detection of small molecules (haptens) by dynamic light scattering biosensing is presented. The proposed technique implements a homogeneous competitive immunoassay and is based on optical detection of specific inhibition of nanoparticle aggregation induced by the analyte in a sample. The technique performance was tested both in buffer and milk for detection of chloramphenicol – antibiotic relevant to food safety diagnostics. Good specificity, sensitivity (LOD in milk is 2.4 ng/ml), precision (4.0 ± 1.2%), ruggedness (8.3%), and 96% recovery in conjunction with a record wide dynamic range (3 orders of magnitude) of the nanosensing technique were demonstrated. Such characteristics complemented by the assay simplicity (no washing step) and a short assay time make the approach attractive for application as an analytical platform for point-of-care and field-oriented diagnostics. Graphical abstractKeywords
Funding Information
- Russian Science Foundation (16-19-00131)
- Ministry of science and higher education of Russian Federation (14.624.21.0045)
This publication has 43 references indexed in Scilit:
- Developing a nanoparticle test for prostate cancer scoringJournal of Translational Medicine, 2012
- Three-dimensional sensing membrane functionalized quartz crystal microbalance biosensor for chloramphenicol detection in real timeSensors and Actuators B: Chemical, 2011
- One-step assay for detecting influenza virus using dynamic light scattering and gold nanoparticlesThe Analyst, 2011
- An ultrasensitive method for the detection of gene fragment from transgenics using label-free gold nanoparticle probe and dynamic light scatteringAnalytica Chimica Acta, 2011
- A sensitive immunoassay based on direct hapten coated format and biotin–streptavidin system for the detection of chloramphenicolTalanta, 2010
- Use of Gold Nanoparticles in a Simple Colorimetric and Ultrasensitive Dynamic Light Scattering Assay: Selective Detection of Arsenic in GroundwaterAngewandte Chemie International Edition, 2009
- A One-Step Homogeneous Immunoassay for Cancer Biomarker Detection Using Gold Nanoparticle Probes Coupled with Dynamic Light ScatteringJournal of the American Chemical Society, 2008
- Optical BiosensorsChemical Reviews, 2008
- One‐Step Homogeneous Detection of DNA Hybridization with Gold Nanoparticle Probes by Using a Linear Light‐Scattering TechniqueAngewandte Chemie International Edition, 2006
- Coagulation Rate Measurements of Colloidal Particles by Simultaneous Static and Dynamic Light ScatteringLangmuir, 1996