Bailout Retrieval of an Entrapped, Fractured, and Undeflatable Stent Balloon in the Left Main Coronary Artery: A Salvage Procedure During Cardiac Arrest
- 30 December 2025
- journal article
- case report
- Published by Wiley in Catheterization and Cardiovascular Interventions
- Vol. 107 (4) , 1120-1126
Abstract
Background: Balloon catheter fracture is a rare PCI complication that may trigger catastrophic coronary entrapment. This represents the first reported cardiac arrest during percutaneous retrieval of an entrapped, fractured stent balloon in an unprotected left main artery. Case Summary: A 72‐year‐old male developed sudden hemodynamic collapse during balloon extraction due to a shaft fracture. Under ongoing CPR with IABP/intubation support, modified balloon anchoring via high‐pressure secondary inflation in 7 F extension catheter enabled en bloc retrieval, with cardiac surgical/ECMO teams preemptively mobilized. Coronary flow was immediately restored, permitting extubation within 1 h and discharge after 5 days without sequelae. Conclusion: Salvage of such events mandates operator agility for bailout adaptation and on‐site availability of a multidisciplinary rescue team.Keywords
This publication has 13 references indexed in Scilit:
- Retrieval of an Undeflatable Balloon During Percutaneous Coronary Intervention in Combination With Small ThoracotomyJACC: Case Reports, 2025
- Clinical expert consensus document on bailout algorithms for complications in percutaneous coronary intervention from the Japanese Association of Cardiovascular Intervention and TherapeuticsCardiovascular Intervention and Therapeutics, 2024
- A rare case report of the successful withdrawal of a stent balloon that failed to deflateBMC Cardiovascular Disorders, 2023
- 2021 ACC/AHA/SCAI Guideline for Coronary Artery RevascularizationJACC, 2022
- A New Technique to Puncture and Remove a PTCA Balloon that Fails to DeflateInternational Journal of Angiology, 2020
- The buddy balloon technique to release a sticky ABSORB bioresorbable vascular scaffold balloon post-scaffoldingInternational Journal of Cardiology, 2016
- Retrieval of impacted broken balloon by balloon inflation in guiding catheterCardiovascular Intervention and Therapeutics, 2013
- Entrapped coronary angioplasty stent balloon due to nondeflation: Percutaneous retrieval by a simple techniqueCatheterization and Cardiovascular Interventions, 2010
- Intracoronary fracture and embolization of a coronary angioplasty balloon catheter: Retrieval by a simple techniqueCatheterization and Cardiovascular Interventions, 2003
- Inflated balloon entrapped in a calcified coronary stenosisCatheterization and Cardiovascular Diagnosis, 1993