Bailout Retrieval of an Entrapped, Fractured, and Undeflatable Stent Balloon in the Left Main Coronary Artery: A Salvage Procedure During Cardiac Arrest

Authors:

Abstract
Background: Balloon catheter fracture is a rare PCI complication that may trigger catastrophic coronary entrapment. This represents the first reported cardiac arrest during percutaneous retrieval of an entrapped, fractured stent balloon in an unprotected left main artery. Case Summary: A 72‐year‐old male developed sudden hemodynamic collapse during balloon extraction due to a shaft fracture. Under ongoing CPR with IABP/intubation support, modified balloon anchoring via high‐pressure secondary inflation in 7 F extension catheter enabled en bloc retrieval, with cardiac surgical/ECMO teams preemptively mobilized. Coronary flow was immediately restored, permitting extubation within 1 h and discharge after 5 days without sequelae. Conclusion: Salvage of such events mandates operator agility for bailout adaptation and on‐site availability of a multidisciplinary rescue team.