Vascular Responses at Proximal and Distal Edges of Paclitaxel-Eluting Stents
- 10 February 2004
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 109 (5) , 627-633
Abstract
Background— On the basis of brachytherapy experience, edge stenosis has been raised as a potential limitation for drug-eluting stents. We used serial intravascular ultrasound (IVUS) to prospectively analyze vessel responses in adjacent reference segments after implantation of polymer-controlled paclitaxel-eluting stents. Methods and Results— TAXUS II was a randomized, double-blind trial with 2 consecutive patient cohorts that compared slow-release (SR) and moderate-release (MR) paclitaxel-eluting stents with control bare metal stents (BMS). By protocol, all patients had postprocedure and 6-month follow-up IVUS. Quantitative IVUS analysis was performed by an independent core laboratory, blinded to treatment allocation, in 5-mm vessel segments immediately proximal and distal to the stent. Serial IVUS was available for 106 SR, 107 MR, and 214 BMS patients. For all 3 groups, a significant decrease in proximal-edge lumen area was observed at 6 months. The decrease was comparable (by ANOVA, P =0.194) for patients in the SR (−0.54±2.1 mm 2 ) and MR (−0.88±1.9 mm 2 ) groups compared with the BMS (−1.02±1.9 mm 2 ) group. For the distal edge, a significant decrease in lumen area was only observed with BMS (−0.91±2.0 mm 2 , P 2 ) and MR (−0.19±1.7 mm 2 ) stents ( P 2 , P =0.011) but not the distal edges of BMS and at neither edge of SR or MR stents. Conclusions— The marked reduction in in-stent restenosis with SR or MR stents is not associated with increased edge stenosis at 6-month follow-up IVUS. In fact, compared with BMS, there is instead a significant reduction in late lumen loss at the distal edge with TAXUS stents.Keywords
This publication has 15 references indexed in Scilit:
- Randomized Study to Assess the Effectiveness of Slow- and Moderate-Release Polymer-Based Paclitaxel-Eluting Stents for Coronary Artery LesionsCirculation, 2003
- TAXUS ICirculation, 2003
- Persistent Inhibition of Neointimal Hyperplasia After Sirolimus-Eluting Stent ImplantationCirculation, 2002
- A Randomized Comparison of a Sirolimus-Eluting Stent with a Standard Stent for Coronary RevascularizationNew England Journal of Medicine, 2002
- Use of localised intracoronary β radiation in treatment of in-stent restenosis: the INHIBIT randomised controlled trialThe Lancet, 2002
- Perivascular Responses after Angioplasty Which May Contribute to Postangioplasty RestenosisAnnals of the New York Academy of Sciences, 2001
- The Biocompatibles’ Drug-Eluting Stents2001
- The TRAPIST Study. A multicentre randomized placebo controlled clinical trial of trapidil for prevention of restenosis after coronary stenting, measured by 3-D intravascular ultrasoundEuropean Heart Journal, 2001
- Localized Intracoronary Gamma-Radiation Therapy to Inhibit the Recurrence of Restenosis after StentingNew England Journal of Medicine, 2001
- I Like the Candy, I Hate the WrapperCirculation, 2000