The long-term mechanical integrity of non-reinforced PEEK-OPTIMA polymer for demanding spinal applications: experimental and finite-element analysis
- 7 June 2005
- journal article
- Published by Springer Nature in European Spine Journal
- Vol. 15 (2) , 149-156
Abstract
Polyetheretherketone (PEEK) is a novel polymer with potential advantages for its use in demanding orthopaedic applications (e.g. intervertebral cages). However, the influence of a physiological environment on the mechanical stability of PEEK has not been reported. Furthermore, the suitability of the polymer for use in highly stressed spinal implants such as intervertebral cages has not been investigated. Therefore, a combined experimental and analytical study was performed to address these open questions. A quasi-static mechanical compression test was performed to compare the initial mechanical properties of PEEK-OPTIMA polymer in a dry, room-temperature and in an aqueous, 37°C environment (n=10 per group). The creep behaviour of cylindrical PEEK polymer specimens (n=6) was measured in a simulated physiological environment at an applied stress level of 10 MPa for a loading duration of 2000 hours (12 weeks). To compare the biomechanical performance of different intervertebral cage types made from PEEK and titanium under complex loading conditions, a three-dimensional finite element model of a functional spinal unit was created. The elastic modulus of PEEK polymer specimens in a physiological environment was 1.8% lower than that of specimens tested at dry, room temperature conditions (P<0.001). The results from the creep test showed an average creep strain of less than 0.1% after 2000 hours of loading. The finite element analysis demonstrated high strain and stress concentrations at the bone/implant interface, emphasizing the importance of cage geometry for load distribution. The stress and strain maxima in the implants were well below the material strength limits of PEEK. In summary, the experimental results verified the mechanical stability of the PEEK-OPTIMA polymer in a simulated physiological environment, and over extended loading periods. Finite element analysis supported the use of PEEK-OPTIMA for load-bearing intervertebral implants.Keywords
This publication has 35 references indexed in Scilit:
- Tensile properties, tension–tension fatigue and biological response of polyetheretherketone–hydroxyapatite composites for load-bearing orthopedic implantsBiomaterials, 2003
- Factors influencing stresses in the lumbar spine after the insertion of intervertebral cages: finite element analysisEuropean Spine Journal, 2002
- Preliminary Experience Using a Polyetheretherketone (PEEK) Cage in the Treatment of Cervical Disc DiseaseNeurosurgery, 2002
- Cervical Cage Fusion with 5 Different Implants: 250 CasesActa Neurochirurgica, 2002
- Resistance of the Lumbar Spine Against Axial Compression Forces after Implantation of Three Different Posterior Lumbar Interbody CagesActa Neurochirurgica, 2001
- Die dorsale lumbale interkorporelle Fusion mit Cages (PLIF) und transpedikulärer StabilisierungCentral European Neurosurgery, 2001
- Attachment and proliferation of osteoblasts and fibroblasts on biomaterials for orthopaedic useBiomaterials, 1995
- Long‐term durability of the interface in FRP composites after exposure to simulated physiologic saline environmentsJournal of Biomedical Materials Research, 1994
- Biological response to chopped‐carbon‐fiber‐reinforced peekJournal of Biomedical Materials Research, 1992
- In vitro biocompatibility of polyetheretherketone and polysulfone compositesJournal of Biomedical Materials Research, 1990