A novel method for determining calibration and behavior of PVDF ultrasonic hydrophone probes in the frequency range up to 100 MHz
Authors:
- P.A. LewinPLP.A. Lewin
- 1 November 2000
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 47 (6) , 1354-1362
Abstract
A new calibration technique for PVDF ultrasonic hydrophone probes is described. Current implementation of the technique allows determination of hydrophone frequency response between 2 and 100 MHz and is based on the comparison of theoretically predicted and experimentally determined pressure-time waveforms produced by a focused, circular source. The simulation model was derived from the time domain algorithm that solves the non linear KZK (Khokhlov-Zabolotskaya-Kuznetsov) equation describing acoustic wave propagation. The calibration technique data were experimentally verified using independent calibration procedures in the frequency range from 2 to 40 MHz using a combined time delay spectrometry and reciprocity approach or calibration data provided by the National Physical Laboratory (NPL), UK. The results of verification indicated good agreement between the results obtained using KZK and the above-mentioned independent calibration techniques from 2 to 40 MHz, with the maximum discrepancy of 18% at 30 MHz. The frequency responses obtained using different hydrophone designs, including several membrane and needle probes, are presented, and it is shown that the technique developed provides a desirable tool for independent verification of primary calibration techniques such as those based on optical interferometry. Fundamental limitations of the presented calibration method are also examined.Keywords
This publication has 26 references indexed in Scilit:
- Nonlinear distortion of short pulses radiated by plane and focused circular pistonsThe Journal of the Acoustical Society of America, 1997
- High-frequency membrane hydrophoneIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1996
- Frequency response of PVDF needle-type hydrophonesUltrasound in Medicine & Biology, 1994
- An experimental investigation of the nonlinear pressure field produced by a plane circular pistonThe Journal of the Acoustical Society of America, 1993
- Distortion and high-frequency generation due to nonlinear propagation of short ultrasonic pulses from a plane circular pistonThe Journal of the Acoustical Society of America, 1992
- Nonlinear pressure fields due to focused circular aperturesThe Journal of the Acoustical Society of America, 1992
- The nonlinear pressure field of a plane circular piston: Theory and experimentThe Journal of the Acoustical Society of America, 1988
- Calibration of hydrophones based on reciprocity and time delay spectrometryIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1988
- Primary calibration of ultrasonic hydrophone using optical interferometryIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1988
- A computerized system for measuring the acoustic output from diagnostic ultrasound equipmentIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1988