Initial Clinical Experience With the Modified Zenith “Pro-Form” TX2 Thoracic Endograft
- 1 August 2010
- journal article
- Published by International Society of Endovascular Specialists in Journal of Endovascular Therapy
- Vol. 17 (4) , 463-470
Abstract
Purpose: To assess the safety and efficacy of thoracic aortic endovascular repair (TEVAR) with the newly introduced Zenith TX2 thoracic stent-graft with Pro-Form. Methods: From March 2009 to present, the Zenith TX2 with Pro-Form was used to treat 27 patients (23 men; median age 65 years, range 22–77) suffering from descending thoracic aortic disease: 20 degenerative aneurysms, 3 chronic type B dissections, 2 type I endoleaks following previous endograft repair, 1 aortic ulcer, and 1 traumatic isthmic rupture. According to Ishimaru's classification, the distribution of the proximal landing zone was zone 0 in 4 cases, zone 1 in 2 cases, zone 2 in 9 cases, zone 3 in 7 cases, and zone 4 in 5 cases. One aneurysm distally involved the thoracoabdominal aorta. An acutely angulated (Results: The device was safely and effectively deployed in all cases, with no problems during sheath retrieval. The postoperative course was free from major complications in all cases, and there was no perioperative mortality. Postoperative morbidity included respiratory failure in 1 case, transient acute renal failure in 3 cases, and post-implantation syndrome in 2 cases. Postoperative computed tomography at 1 and 6 months confirmed 100% clinical success with the absence of device-related complications. No type I endoleak was documented. The proximal sealing stent fitted well with the aortic profile in all cases except one, in which malapposition was seen in a severely angulated aorta at 6-month follow-up. Conclusion: Based on this preliminary experience, the TX2 thoracic stent-graft with Pro-Form can be safely and effectively deployed to satisfactorily repair thoracic aortic pathology, including lesions in the arch. Stent-graft conformance to the aorta was excellent even in cases with unfavorable anatomy. These encouraging results will need to be confirmed in larger series with larger follow-up.Keywords
This publication has 23 references indexed in Scilit:
- Factors favoring stent-graft collapse after thoracic endovascular aortic repairThe Journal of Thoracic and Cardiovascular Surgery, 2010
- Infolding and collapse of thoracic endoprostheses: Manifestations and treatment optionsThe Journal of Thoracic and Cardiovascular Surgery, 2009
- Magnitude and Direction of Pulsatile Displacement Forces Acting on Thoracic Aortic EndograftsJournal of Endovascular Therapy, 2009
- Endograft Collapse After Thoracic Stent-Graft Repair for Traumatic RuptureThe Annals of Thoracic Surgery, 2009
- Failure modes of thoracic endografts: Prevention and managementJournal of Vascular Surgery, 2009
- Thoracic endovascular stent graft with a bird’s beak signCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 2008
- Pivotal results of the Medtronic Vascular Talent Thoracic Stent Graft System: The VALOR TrialJournal of Vascular Surgery, 2008
- International controlled clinical trial of thoracic endovascular aneurysm repair with the Zenith TX2 endovascular graft: 1-year resultsJournal of Vascular Surgery, 2008
- Endovascular treatment of thoracic aortic aneurysms: Results of the phase II multicenter trial of the GORE TAG thoracic endoprosthesisJournal of Vascular Surgery, 2005
- Disappointing results with a new commercially available thoracic endograftJournal of Vascular Surgery, 2004