Comparing iStent complications across four generations: a manufacturer and user facility device experience (MAUDE) database study
- 12 November 2025
- journal article
- abstracts
- Published by Springer Nature in Graefe's Archive for Clinical and Experimental Ophthalmology
- Vol. 264 (3) , 795-803
Abstract
Purpose To report real-world intraoperative and early postoperative (< 3 months) complications of iStent surgery and analyze notable examples. Patients and methods The Manufacturer and User Facility Device Experience database was queried for “iStent” reports from January 2014 to March 2024. Late postoperative events, reports containing inadequate detail, and those from indirect sources (e.g., published literature, social media posts) were excluded. Adverse events were analyzed per device generation. Results A total of 808 reports were included, revealing 437, 377, 465, and 48 adverse events across the different iStent generations. The most common intraoperative events were malposition for iStent Trabecular Micro-Bypass, Inject, and Infinite (35%, 40%, 35%) and trocar damage for Inject W (34%). Surgeons and the manufacturer most frequently cited trocar bias as the cause of trocar damage. When the timing of trocar damage was specified, 66.7% occurred during deployment of the final stent of the series. Notably, device-derived foreign body contamination was exclusive to Inject W (11). The most common early postoperative event for all four generations was elevated intraocular pressure (IOP), occurring in 32%, 38%, 15%, and 15% of cases respectively. For the Infinite, hyphema was observed at a comparable proportion (15%). Conclusion Post-market data highlight distinct complication profiles across iStent generations, with malposition and elevated IOP as leading reported issues. Generation-specific trends, such as trocar damage in Inject W and Infinite, underscore the need for tailored surgical vigilance and further study.Keywords
Funding Information
- Research to Prevent Blindness (Challenge Grant)
This publication has 15 references indexed in Scilit:
- Adverse Events Associated With 3 Vitrectomy Platforms Reported to the US FDA MAUDE DatabaseJournal of Vitreoretinal Diseases, 2024
- iStent insertion orientation and impact on trabecular meshwork motion resolved by optical coherence tomography imagingJournal of Biomedical Optics, 2024
- Intraocular Inflammation with faricimab: insights from Manufacturer and User Facility Device Experience (MAUDE) databaseEye, 2024
- A case of two connected stents deployed during iStent inject W surgeryBMC Ophthalmology, 2023
- Postoperative Endophthalmitis after Combined Cataract Extraction and iStent Inject ImplantationCase Reports in Ophthalmological Medicine, 2023
- Endophthalmitis after Minimally Invasive Glaucoma SurgeryOphthalmology, 2021
- A systematic review of the quality of cardiovascular surgery studies that extracted data from the MAUDE databaseJournal of Vascular Surgery, 2021
- Adverse events associated with femtosecond laser–assisted cataract surgery reported to the MAUDE databaseJournal of Cataract & Refractive Surgery, 2021
- Adverse Events Associated with Microinvasive Glaucoma Surgery Reported to the Food and Drug AdministrationOphthalmology Glaucoma, 2021
- Barriers to Reporting: Limitations of the Maude DatabaseDermatologic Surgery, 2021