Detecting high‐frequency oscillations in real time during epilepsy surgery with neuromorphic hardware validated to predict postoperative seizure outcome
Authors:
- Jeroen TeurlingsJTJeroen Teurlings
- Filippo CostaFCFilippo Costa
- Maeike ZijlmansMZMaeike Zijlmans
- Maryse A. van ‘t KloosterMKMaryse A. van ‘t Klooster
- Giacomo IndiveriGIGiacomo Indiveri
- Niklaus KrayenbühlNKNiklaus Krayenbühl
- Murezi CapaulMCMurezi Capaul
- Debora LedergerberDLDebora Ledergerber
- Georgia RamantaniGRGeorgia Ramantani
- Johannes SarntheinJSJohannes Sarnthein
Open Access
Abstract
Objective: High‐frequency oscillations (HFOs; 250–500 Hz) in electrocorticography (ECoG) are promising biomarkers for delineating the epileptogenic zone during epilepsy surgery. Accurate, intraoperative real‐time detection of HFOs may guide surgical resection to improve seizure outcome. Methods: We used intraoperative ECoG from 22 patients from Zurich (Nicolet recording device, recording under Sevoflurane) with annotated HFOs to train an innovative automated HFO detector implemented in a spiking neural network (SNN) on neuromorphic hardware. We validated whether the detector could identify clinically relevant HFOs in an independent cohort of 54 patients from Utrecht (Micromed recording device, recording during a pause in propofol administration) by examining whether HFOs remained after resection and whether this predicted seizure outcome. We then tested the detector in real time during surgery, giving direct feedback to the neurosurgeon. Results: The optimal detection parameters yielded Spearman R = .71 between channelwise HFO rates detected by the proposed SNN detection and those previously detected by the Spectrum Detector in the training dataset. We found postresection HFOs in nine patients who had poor surgery outcome and in one patient with good surgery outcome in the validation data, resulting in a diagnostic odds ratio = 17 (95% confidence interval = 2.0–150). During a live surgery, the detector analyzed incoming intraoperative ECoG during acquisition, HFOs were successfully captured, and HFO feedback was available within 1 min after completion of the recording period. Significance: Our automatic detector was easily transferred between centers and anesthesia protocols. We showed the direct use of intraoperative HFO detection as possible real‐time feedback to the surgeon to tailor resection, yielding high predictive power of seizure outcome.Keywords
Funding Information
- European Research Council (803880)
- Anna Mueller Grocholski-Stiftung
- ZonMw (091501722100587)
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung (10008856)
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung (204651)
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung (208184)
This publication has 21 references indexed in Scilit:
- Neuromorphic is dead. Long live neuromorphic.Neuron, 2025
- Using high-frequency oscillations from brief intraoperative neural recordings to predict the seizure onset zoneCommunications Medicine, 2024
- Normative atlases of high-frequency oscillation and spike rates under Sevoflurane anesthesiaClinical Neurophysiology, 2024
- Pausing propofol during neurosurgery to record intraoperative electrocorticography is feasible;10 years of clinical experienceClinical Neurophysiology, 2024
- Robust compression and detection of epileptiform patterns in ECoG using a real-time spiking neural network hardware frameworkNature Communications, 2024
- Intraoperative electrocorticography using high-frequency oscillations or spikes to tailor epilepsy surgery in the Netherlands (the HFO trial): a randomised, single-blind, adaptive non-inferiority trialThe Lancet Neurology, 2022
- An electronic neuromorphic system for real-time detection of high frequency oscillations (HFO) in intracranial EEGNature Communications, 2021
- A spiking neural network (SNN) for detecting high frequency oscillations (HFOs) in the intraoperative ECoGScientific Reports, 2021
- High-density ECoG improves the detection of high frequency oscillations that predict seizure outcomeClinical Neurophysiology, 2019
- Automatic detection of high frequency oscillations during epilepsy surgery predicts seizure outcomeClinical Neurophysiology, 2016