Abstract
Silver(I) selective sensors have been fabricated from polystyrene matrix membranes containing macrocycle, Me6(14) diene.2HClO4 as ionophore. Best performance was exhibited by the membrane having a composition macrocycle : Polystyrene in the ratio 15:1. This membrane worked well over a wide concentration range 5.0×10-6–1.0×10-1M of Ag+ with a near-Nernstian slope of 53.0 ± 1.0 mV per decade of Ag+ activity. The response time of the sensor is + in solution having up to 25% (v/v) nonaqueous fraction.

This publication has 17 references indexed in Scilit: