Evaluating the Clinical Accuracy of Two Continuous Glucose Sensors Using Continuous Glucose–Error Grid Analysis
Open Access
- 1 October 2005
- journal article
- clinical trial
- Published by American Diabetes Association in Diabetes Care
- Vol. 28 (10) , 2412-2417
Abstract
OBJECTIVE—To compare the clinical accuracy of two different continuous glucose sensors (CGS) during euglycemia and hypoglycemia using continuous glucose–error grid analysis (CG-EGA). RESEARCH DESIGN AND METHODS—FreeStyle Navigator (Abbott Laboratories, Alameda, CA) and MiniMed CGMS (Medtronic, Northridge, CA) CGSs were applied to the abdomens of 16 type 1 diabetic subjects (age 42 ± 3 years) 12 h before the initiation of the study. Each system was calibrated according to the manufacturer’s recommendations. Each subject underwent a hyperinsulinemic-euglycemic clamp (blood glucose goal 110 mg/dl) for 70–210 min followed by a 1–mg · dl−1 · min−1 controlled reduction in blood glucose toward a nadir of 40 mg/dl. Arterialized blood glucose was determined every 5 min using a Beckman Glucose Analyzer (Fullerton, CA). CGS glucose recordings were matched to the reference blood glucose with 30-s precision, and rates of glucose change were calculated for 5-min intervals. CG-EGA was used to quantify the clinical accuracy of both systems by estimating combined point and rate accuracy of each system in the euglycemic (70−180 mg/dl) and hypoglycemic (r = 0.84; CGMS, r = 0.79, NS). During euglycemia, both CGS systems had similar clinical accuracy (Navigator zones A + B, 88.8%; CGMS zones A + B, 89.3%, NS). However, during hypoglycemia, the Navigator was significantly more clinically accurate than the CGMS (zones A + B = 82.4 vs. 61.6%, Navigator and CGMS, respectively, P < 0.0005). CONCLUSIONS—CG-EGA is a helpful tool for evaluating and comparing the clinical accuracy of CGS systems in different blood glucose ranges. CG-EGA provides accuracy details beyond other methods of evaluation, including correlational analysis and the original EGA.Keywords
This publication has 9 references indexed in Scilit:
- Evaluating the Accuracy of Continuous Glucose-Monitoring SensorsDiabetes Care, 2004
- Accuracy of the GlucoWatch G2 Biographer and the Continuous Glucose Monitoring System During HypoglycemiaDiabetes Care, 2004
- The Need for Separate Performance Goals for Glucose Sensors in the Hypoglycemic, Normoglycemic, and Hyperglycemic RangesDiabetes Care, 2004
- The Accuracy of the CGMS ™ in Children with Type 1 Diabetes: Results of the Diabetes Research in Children Network (DirecNet) Accuracy StudyDiabetes Technology & Therapeutics, 2003
- Performance Evaluation of Blood Glucose Monitoring DevicesDiabetes Technology & Therapeutics, 2003
- A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose.Diabetes Care, 2000
- Newer Portable Glucose Meters—Analytical Improvement Compared with Previous Generation Devices?Clinical Chemistry, 1999
- Validation of Home Blood Glucose Meters With Respect to Clinical and Analytical ApproachesDiabetes Care, 1998
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987