Improved k‐space trajectory measurement with signal shifting
Open Access
- 20 July 2007
- journal article
- other
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 58 (1) , 200-205
Abstract
Various techniques for k‐space trajectory measurement have been described in the literature. Self‐encoding gradient techniques are time‐consuming due to the high number of phase‐encoding steps needed. The approach with localized slices is faster, but its use apparently has not been reported in the context of high spatial resolution experiments. Signals associated with high k‐space frequencies may then reach low or even zero values, and this may result in errors in the estimate of the trajectories at the k‐space periphery. To overcome this problem without increasing the measurement duration of the localized slice method too much, a new approach is proposed in which an addition dephasing gradient applied prior to the gradient to be measured shifts the signal maximum. Magn Reson Med 58:200–205, 2007.Keywords
This publication has 8 references indexed in Scilit:
- Spiral imaging: A critical appraisalJournal of Magnetic Resonance Imaging, 2005
- A novel k‐space trajectory measurement techniqueMagnetic Resonance in Medicine, 1998
- Simple Correction Method fork-Space Trajectory Deviations in MRIJournal of Magnetic Resonance, 1998
- Gradient characterization using a Fourier‐transform techniqueMagnetic Resonance in Medicine, 1998
- A Method to measure arbitrary k‐space trajectories for rapid MR imagingMagnetic Resonance in Medicine, 1997
- An integrated program for amplitude-modulated RF pulse generation and re-mapping with shaped gradientsMagnetic Resonance Imaging, 1994
- Analysis of the eddy-current induced artifacts and the temporal compensation in nuclear magnetic resonance imagingIEEE Transactions on Medical Imaging, 1991
- The k‐trajectory formulation of the NMR imaging process with applications in analysis and synthesis of imaging methodsMedical Physics, 1983