A programmable electronic stimulator for FES systems
Authors:
- 1 December 1994
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Rehabilitation Engineering
- Vol. 2 (4) , 234-239
Abstract
A low-cost, programmable, galvanically isolated bipolar four channel stimulator for patterned electrical stimulation has been developed. The design is based on a newly developed concept, and it includes a high efficiency DC/DC converter securing high output voltage and sufficient energy to drive four output stages, a microcontroller for control of charge compensation, pulse rate and duration, as well as patterns of exercise and functional use regimes. A user-friendly interface for programming of the stimulator has been developed for the PC host computer, using Windows environment and a push-button control for functional application of the system. The use in humans who suffered a central nervous system lesion and lost voluntary control over muscles necessary for standing, walking, or manipulation was tested. The prototype is small, light, battery-operated, and flexible for various applications.< >Keywords
This publication has 4 references indexed in Scilit:
- Control aspects of active above-knee prosthesisInternational Journal of Man-Machine Studies, 1991
- A flexible, portable system for neuromuscular stimulation in the paralyzed upper extremityIEEE Transactions on Biomedical Engineering, 1988
- An Externally Powered, Multichannel, Implantable Stimulator for Versatile Control of Paralyzed MuscleIEEE Transactions on Biomedical Engineering, 1987
- Electrical splinting of the knee in paraplegiaSpinal Cord, 1979