Cross-sensitivities of electrochemical detectors used to monitor worker exposures to airborne contaminants: False positive responses in the absence of target analytes
- 23 November 2005
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Environmental Monitoring
- Vol. 8 (1) , 161-166
Abstract
Ideally, the response of electrochemical detectors is proportional to the concentration of targeted airborne chemicals and is not be affected by concomitantly present substances. Manufacturers provide a limited list of cross-sensitivities but end-users have anecdotally reported unexpected interferences by other substances. Electrochemical detectors designed to measure airborne levels of CO, H2S, NO, NO2, or SO2, were challenged with potentially interfering substances in the absence of target analytes. Cross-sensitivities undocumented by the manufacturers were observed and were found to vary between different models of instruments for the same challenge chemical.Keywords