Effects of Prolonged Ambient Storage of Sodium Fluoride/Heparin Specimens on Plasma Homocysteine
Open Access
- 1 August 2005
- journal article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 51 (8) , 1564-1565
Abstract
Increased plasma homocysteine is regarded as a risk factor for arterial and venous occlusive disease. Sample collection ideally requires use of cooled EDTA containers and immediate centrifugation of the blood (1). These conditions may be hard to maintain in a clinic and are not possible in the many medical practices that lack a centrifuge. We tested whether analyzing uncentrifuged EDTA or NaF/heparin plasma bears a greater risk of giving false-positive homocysteine results under these conditions. Blood was drawn from 50 healthy volunteers (40 female and 10 male) into 1 EDTA and 4 NaF/heparin Vacutainers (Becton Dickinson), which were stored at room temperature. The Vacutainers were randomly assigned to groups—EDTA-15min, NaF/heparin-15min, NaF/heparin-24h, NaF/heparin-48h, and NaF/heparin-144h—and centrifuged at 2700 g for …Keywords
This publication has 4 references indexed in Scilit:
- Total Plasma Homocysteine Measured by Liquid Chromatography–Tandem Mass Spectrometry with Use of 96-Well PlatesClinical Chemistry, 2004
- Facts and Recommendations about Total Homocysteine Determinations: An Expert OpinionClinical Chemistry, 2004
- Stabilization of Homocysteine in Unseparated Blood over Several Days: A Solution for Epidemiological StudiesClinical Chemistry, 2003
- Addition of Sodium Fluoride to Whole Blood Does Not Stabilize Plasma Homocysteine But Produces Dilution Effects on Plasma Constituents and HematocritClinical Chemistry, 1998