Determining fitting ranges of various bone conduction hearing aids
- 29 May 2017
- journal article
- research article
- Published by Wiley in Clinical Otolaryngology
- Vol. 43 (1) , 68-75
Abstract
Objectives: To define fitting ranges for nine bone conduction devices (BCDs) over different frequencies based on the device's maximum power output (MPO) and to validate the assessment of MPO of BCDs in the ear canal.Background: Maximum power output (MPO) is an important characteristic when fitting BCDs. It is the highest output level a device can deliver and is one of the major determinants of a device's fitting range. A skull simulator can be used to verify MPO of percutaneous BCDs. No such simulator is available for active and passive transcutaneous devices.Design: The MPO of nine different BCDs was assessed either by real‐ear measurements and/or with skull simulator measurements.Main outcome measures: MPO and cross‐validation of the methods using the Bland–Altman method.Results: Percutaneous BCDs have higher MPO levels compared to active and passive transcutaneous devices. This results in a wide dynamic range of hearing for percutaneous devices. Moreover, the assessment of MPO by real‐ear measurements was validated.Conclusion: Based on MPO data, fitting ranges were defined for nine BCDs over seven frequencies.Keywords
This publication has 15 references indexed in Scilit:
- An Experimental Objective Method to Determine Maximum Output and Dynamic Range of an Active Bone Conduction ImplantOtology & Neurotology, 2014
- Amplification Options for Patients With Mixed Hearing LossOtology & Neurotology, 2014
- Comparison Between a New Implantable Transcutaneous Bone Conductor and Percutaneous Bone-Conduction Hearing ImplantOtology & Neurotology, 2013
- Fitting range of the BAHA CordelleInternational Journal of Audiology, 2006
- Gain and Maximum Output of Two Electromagnetic Middle Ear Implants: Are Real Ear Measurements Helpful?Journal of the American Academy of Audiology, 2004
- The Bone-Anchored Hearing Aid: Reference Quantities and Functional GainEar & Hearing, 1997
- Force threshold for hearing by direct bone conductionThe Journal of the Acoustical Society of America, 1995
- Transfer Functions and Correction Factors Used in Hearing Aid Evaluation and ResearchEar & Hearing, 1989
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Hearing Thresholds with Direct Bone Conduction Versus Conventional Bone ConductionScandinavian Audiology, 1984