Percutaneous balloon atrial septostomy on top of venoarterial extracorporeal membrane oxygenation results in safe and effective left heart decompression
- 14 October 2016
- journal article
- research article
- Published by Oxford University Press (OUP) in European Heart Journal: Acute Cardiovascular Care
- Vol. 7 (1) , 70-79
Abstract
Background: Transcatheter techniques are emerging for left atrial (LA) decompression under venoarterial extracorporeal membrane oxygenation (VA-ECMO). We aimed to assess whether balloon atrioseptostomy (BAS) is a safe and efficient strategy. Methods: All patients who underwent percutaneous static BAS under VA-ECMO at four tertiary institutions were retrospectively reviewed. Results: From 2000 to 2014, BAS was performed in 64 patients (32 adults and 32 children). Indications for ECMO support included acute myocarditis (31.3%) and non-myocarditis cardiac disease, mostly end-stage dilated cardiomyopathy (32.8%). BAS was required because of pulmonary oedema/haemorrhage and left ventricular (LV) distension. The mean balloon diameter was 21.8 ± 8.4mm. Adequate LA decompression was achieved in all patients. Mean LA pressure fell from 24.2 ± 6.9 mmHg to 7.8 ± 2.6 mmHg (p < 0.001). The left-to-right atrial pressure gradient fell from 17.2 ± 7.1 mmHg to 0.09 ± 0.5 mmHg (p < 0.001). Echocardiography showed an unrestrictive left-to-right atrial shunting in all patients. Improvement of day 1 chest X-ray was observed in 76.6% of patients, clinical status in 98.4% of patients and pulmonary haemorrhage in 14 out of 14 patients. Complications occurred in 9.4% of patients, representing pericardial effusion, fast atrial fibrillation, ventricular fibrillation requiring defibrillation, transient complete heart block and femoral venous dissection requiring covered stent placement. In the 37 (57.8%) patients who were successfully decannulated, the median ECMO duration was 9 (range: 4–24) days. After a median follow-up of 12.3 (range: 0.1–142) months, 35.9% patients died, 17.2% received a LV assist device as a bridge to transplantation, 31.2% were transplanted and 56.2% were home discharged and alive. Conclusions: Percutaneous BAS may be a safe and efficient strategy for discharging the LA in both adults and children supported by VA-ECMO.Keywords
This publication has 34 references indexed in Scilit:
- Early catheterization after initiation of extracorporeal membrane oxygenation support in children is associated with improved survivalCatheterization and Cardiovascular Interventions, 2016
- Extracorporeal membrane oxygenation for critically ill adultsCochrane Database of Systematic Reviews, 2015
- Middle‐term results of trans‐catheter creation of atrial communication in patients receiving mechanical circulatory supportCatheterization and Cardiovascular Interventions, 2015
- Outcomes of Pediatric Patients Undergoing Cardiac Catheterization While on Extracorporeal Membrane OxygenationPediatric Cardiology, 2014
- Transseptal Catheter Decompression of the Left Ventricle During Extracorporeal Membrane OxygenationPediatric Cardiology, 2011
- Venoarterial Extracorporeal Membrane Oxygenation (VA-ECMO) in Pediatric Cardiac SupportThe Annals of Thoracic Surgery, 2006
- Blade and balloon atrial septostomy for left heart decompression in patients with severe ventricular dysfunction on extracorporeal membrane oxygenationCatheterization and Cardiovascular Interventions, 1999
- Bedside transseptal balloon dilation atrial septostomy for decompression of the left heart during extracorporeal membrane oxygenationCatheterization and Cardiovascular Interventions, 1999
- Transseptal decompression of the left heart during ECMO for severe myocarditisThe Annals of Thoracic Surgery, 1995
- Echocardiographically guided balloon atrial septostomy during extracorporeal membrane oxygenation (ECMO)Pediatric Cardiology, 1993