Report of Two Cases of Bladder Perforation Caused by Reservoir of Inflatable Penile Prosthesis
- 1 July 2009
- journal article
- case report
- Published by Oxford University Press (OUP) in The Journal of Sexual Medicine
- Vol. 6 (7) , 2064-2067
Abstract
Introduction: Rupture of penile prosthesis reservoir into the bladder is an unusual and unexpected occurrence after prosthesis component replacement for malfunction. Scant literature exists describing a pathophysiology, diagnosis, and repair with device salvage. Aim: Our goal is to describe an uncommon event in prosthetic surgery, and discuss the pathophysiology and ways to treat and perhaps avoid this occurrence. Methods: Two patients presented with gross hematuria in the recovery room following repair of a cylinder to pump tubing break of their inflatable penile prostheses (IPPs). In each operation, the existing reservoir had been left in situ and refilled once placed in circuit with new cylinders and pump. Results: The reservoirs had immediately ruptured into the bladder upon refilling, and were identified intravesically on cystoscopy and computed tomography scan. A one-step approach can be done to salvage the device, remove the eroded reservoir with bladder repair, and new reservoir placed directly into appropriate space. Interpretation: During IPP revision, even when a reservoir is not manipulated simple tubing repair performed, gross hematuria in the recovery room likely signals bladder perforation. We include recommendations for reservoir management with implant revision, as well as surgical management and prevention of this complication. Conclusions: Bladder perforation can be caused by penile implant revision even when the old reservoir is not manipulated. If this complication occurs, it can be surgically managed immediately. This can also potentially be avoided.Keywords
This publication has 10 references indexed in Scilit:
- Revision Washout Decreases Implant Capsule Tissue Culture Positivity: A Multicenter StudyJournal of Urology, 2008
- Rare Complication of Multiple Revision Surgeries of Penile ProsthesisThe Journal of Sexual Medicine, 2005
- Retrieval of inflatable penile prosthesis reservoir from the bladderInternational Journal Of Impotence Research, 2005
- FATE OF THE RETAINED RESERVOIR AFTER REPLACEMENT OF 3-PIECE PENILE PROSTHESISJournal of Urology, 2004
- PENILE PROSTHESIS CULTURES DURING REVISION SURGERY: A MULTICENTER STUDYJournal of Urology, 2004
- Inflatable penile prosthesis: site-specific malfunction analysisInternational Journal Of Impotence Research, 2003
- Safety and Efficacy Outcome of Mentor Alpha-1 Inflatable Penile Prosthesis Implantation for Impotence TreatmentJournal of Urology, 1997
- Erosion of an Inflatable Penile Prosthesis Reservoir into an Ileal ConduitJournal of Urology, 1987
- Erosion of Inflatable Penile Prosthesis Reservoir Into BladderJournal of Urology, 1986
- Erosion of Inflatable Penile Prosthesis Reservoir into Bladder and BowelJournal of Urology, 1984