Accurate scatter correction for transmission computed tomography using an uncollimated line array source
Authors:
- 1 February 2004
- journal article
- research article
- Published by Springer Nature in Annals of Nuclear Medicine
- Vol. 18 (1) , 45-50
Abstract
We investigated scatter correction in transmission computed tomography (TCT) imaging by the combination of an uncollimated transmission source and a parallel-hole collimator. We employed the triple energy window (TEW) as the scatter correction and found that the conventional TEW method, which is accurate in emission computed tomography (ECT) imaging, needs some modification in TCT imaging based on our phantom studies. In this study a Tc-99m uncollimated line array source (area: 55 cm × 40 cm) was attached to one camera head of a dual-head gamma camera as a transmission source, and TCT data were acquired with a low-energy, general purpose (LEGP), parallel-hole collimator equipped on the other camera head. The energy spectra for 140 keV-photons transmitted through various attenuating material thicknesses were measured and analyzed for scatter fraction. The results of the energy spectra showed that the photons transmitted had an energy distribution that constructs a scatter peak within the 140 keV-photopeak energy window. In TCT imaging with a cylindrical water phantom, the conventional TEW method with triangle estimates (subtraction factor,K = 0.5) was not sufficient for accurate scatter correction (μ = 0.131 cm-1 for water), whereas the modified TEW method withK= 1.0 gave the accurate attenuation coefficient of 0.153 cm-1 for water. For the TCT imaging with the combination of the uncollimated Tc-99m line array source and parallel hole collimator, the modified TEW method withK = 1.0 gives the accurate TCT data for quantitative SPECT imaging in comparison with the conventional TEW method withK= 0.5.Keywords
This publication has 12 references indexed in Scilit:
- Quantitative planar imaging method for measurement of renal activity by using a conjugate-emission image and transmission dataMedical Physics, 2000
- Estimation of scattered photons in gamma ray transmission CT using Monte Carlo simulationsIEEE Transactions on Nuclear Science, 1997
- Quantitative analysis of scatter- and attenuation-compensated dynamic single-photon emission tomography for functional hepatic imaging with a receptor-binding radiopharmaceuticalEuropean Journal of Nuclear Medicine and Molecular Imaging, 1997
- Evaluation of SPET quantification of simultaneous emission and transmission imaging of the brain using a multidetector SPET system with the TEW scatter compensation method and fan-beam collimationEuropean Journal of Nuclear Medicine and Molecular Imaging, 1996
- Attenuation compensation for cardiac single-photon emission computed tomographic imaging: Part 2. Attenuation compensation algorithmsJournal of Nuclear Cardiology, 1996
- Multiwindow scatter correction techniques in single‐photon imagingMedical Physics, 1995
- Attenuation compensation for cardiac single-photon emission computed tomographic imaging: Part 1. Impact of attenuation and methods of estimating attenuation mapsJournal of Nuclear Cardiology, 1995
- A transmission‐map‐based scatter correction technique for SPECT in inhomogeneous mediaMedical Physics, 1995
- Effect of energy resolution on scatter fraction in scintigraphic imaging: Monte Carlo studyMedical Physics, 1993
- A practical method for position-dependent Compton-scatter correction in single photon emission CTIEEE Transactions on Medical Imaging, 1991