A New Method to Make 24-Hour Urine Collection More Convenient: A Validity Study
Open Access
- 20 May 2014
- journal article
- research article
- Published by Wiley in International Journal of Nephrology
- Vol. 2014, 1-5
Abstract
Background and Objectives. This study proposes a novel urine collection device that can divide each urine collection into 20 parts and store and cool just one part. The aim of the current study is to compare measured biomarkers from the proposed urine collection device to those of conventional 24-hour sampling method. We also hypothesized that the new method would significantly increase patients’ adherence to the timed urine collection.Methods. Two 24-hour urine samples with the conventional method and with the new automated urine collection device that uses just one-twentieth of each void were obtained from 40 healthy volunteers. Urine parameters including volume, creatinine, and protein levels were compared between the two methods and the agreement of two measurements for each subject was reported through Bland-Altman plots.Results. Our results confirmed that for all three variables, there is a positive correlationbetween the two measurements and high degree of agreement could be seen in Bland-Altman plots. Moreover, more subjects reported the new method as “more convenient” for 24-hour urine collection.Conclusions. Our results clearly indicate that a fixed proportion of each void may significantly reduce the urine volume in timed collections and this, in turn, may increase subjects’ adherence to this difficult sampling.
Keywords
This publication has 14 references indexed in Scilit:
- Low level of adherence to instructions for 24-hour urine collection among hospital outpatientsBiochemia Medica, 2013
- An audit of 24-hour creatinine clearance measurements at Tygerberg Hospital and comparison with prediction equations.2007
- Estimating kidney function in adults using formulaeAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2005
- The Relationship Between Magnitude of Proteinuria Reduction and Risk of End-stage Renal DiseaseArchives of internal medicine (1960), 2005
- A simple method for correcting overestimated glomerular filtration rate in obese subjects evaluated by the Cockcroft and Gault formula: a comparison with 51Cr EDTA clearanceClinical Nephrology, 2004
- A Comparison of Prediction Equations for Estimating Glomerular Filtration Rate in Adults without Kidney DiseaseJournal of the American Society of Nephrology, 2003
- Assessment of glomerular filtration rate in healthy subjects and normoalbuminuric diabetic patients: validity of a new (MDRD) prediction equationNephrology Dialysis Transplantation, 2002
- Methodological issues in studying the epidemiology of mild to moderate chronic renal insufficiencyKidney International, 2002
- Abnormal Laboratory Results: Creatinine clearance and the assessment of renal functionAustralian Prescriber, 2001
- Blood Pressure Control, Proteinuria, and the Progression of Renal DiseaseAnnals of Internal Medicine, 1995