Calibration of an amorphous‐silicon flat panel portal imager for exit‐beam dosimetry
Authors:
- Cynthia F. ChuangCCCynthia F. Chuang
- Olivier MorinOMOlivier Morin
- Michèle AubinMAMichèle Aubin
- Jean PouliotJPJean Pouliot
- 9 February 2006
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 33 (3) , 584-594
Abstract
Amorphous‐silicon flat panel detectors are currently used to acquire digital portal images with excellent image quality for patient alignment before external beam radiation therapy. As a first step towards interpreting portal images acquired during treatment in terms of the actual dose delivered to the patient, a calibration method is developed to convert flat panel portal images to the equivalent water dose deposited in the detector plane and at a depth of . The method is based on empirical convolution models of dose deposition in the flat panel detector and in water. A series of calibration experiments comparing the response of the flat panel imager and ion chamber measurements of dose in water determines the model parameters. Kernels derived from field size measurements account for the differences in the production and detection of scattered radiation in the two systems. The dissimilar response as a function of beam energy spectrum is characterized from measurements performed at various off‐axis positions and for increasing attenuator thickness in the beam. The flat panel pixel inhomogeneity is corrected by comparing a large open field image with profiles measured in water. To verify the accuracy of the calibration method, calibrated flat panel profiles were compared with measured dose profiles for fields delivered through solid water slabs, a solid water phantom containing an air cavity, and an anthropomorphic head phantom. Open rectangular fields of various sizes and locations as well as a multileaf collimator‐shaped field were delivered. For all but the smallest field centered about the central axis, the calibrated flat panel profiles matched the measured dose profiles with little or no systematic deviation and approximately 3% (two standard deviations) accuracy for the in‐field region. The calibrated flat panel profiles for fields located off the central axis showed a small systematic deviation from the measured profiles for the in‐field region. Out of the field, the differences between the calibrated flat panel and measured profiles continued to be small, approximately 0%–2% of the mean in‐field dose. Further refinement of the calibration model should increase the accuracy of the procedure. This calibration method for flat panel portal imagers may be used as part of a validation scheme to verify the dose delivered to the patient during treatment.Keywords
This publication has 32 references indexed in Scilit:
- The long‐term stability of amorphous silicon flat panel imaging devices for dosimetry purposesMedical Physics, 2004
- Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapyMedical Physics, 2003
- Blurring artifacts in megavoltage radiography with a flat‐panel imaging system: Comparison of Monte Carlo simulations with measurementsMedical Physics, 2002
- An iterative EPID calibration procedure for dosimetric verification that considers the EPID scattering factorMedical Physics, 2001
- Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging deviceMedical Physics, 2001
- Dosimetric verification of intensity modulated beams produced with dynamic multileaf collimation using an electronic portal imaging deviceMedical Physics, 1999
- Relative dosimetry using active matrix flat-panel imager (AMFPI) technologyMedical Physics, 1999
- Transmission dosimetry with a liquid-filled electronic portal imaging deviceInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1996
- Portal dose measurement in radiotherapy using an electronic portal imaging device (EPID)Physics in Medicine & Biology, 1995
- The use of an electronic portal imaging device for exit dosimetry and quality control measurementsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1995