Ultrahigh dose-rate (FLASH) x-ray irradiator for pre-clinical laboratory research
Authors:
- 23 April 2021
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 66 (9) , 095006
Abstract
FLASH irradiation has been shown to reduce significantly normal tissue toxicity compared to conventional irradiation, while maintaining tumor control probability at similar level. Clinical translation of FLASH irradiation necessitates comprehensive laboratory studies to elucidate biological effects as well as pertinent technological and physical requirements. At present, FLASH research employs complex accelerator technologies of limited accessibilities. Here, we study the feasibility of a novel self-shielded x-ray irradiation cabinet system, as an enabling technology to enhance the preclinical research capabilities. The proposed system employs two commercially available high capacity 150 kVp fluoroscopy x-ray sources with rotating anode technology in a parallel-opposed arrangement. Simulation was performed with the GEANT4 Monte-Carlo platform. Simulated dosimetric properties of the x-ray beam for both FLASH and conventional dose-rate irradiations were characterized. Dose and dose rate from a single kV x-ray fluoroscopy source in solid water phantom were verified with measurements using Gafchromic films. The parallel-opposed x-ray sources can deliver over 50 Gy doses to a 20 mm thick water equivalent medium at ultrahigh dose-rates of 40–240 Gy s −1 . A uniform depth-dose rate (±5%) is achieved over 8–12 mm in the central region of the phantom. Mirrored beams minimize heel effect of the source and achieve reasonable cross-beam uniformity (±3%). Conventional dose-rate irradiation (≤0.1 Gy s −1 ) can also be achieved by reducing the tube current and increasing the distance between the phantom and tubes. The rotating anode x-ray source can be used to deliver both FLASH and conventional dose-rate irradiations with the field dimensions well suitable for small animal and cell-culture irradiations. For FLASH irradiation using parallel-opposed sources, entrance and exit doses can be higher by 30% than the dose at the phantom center. Beam angling can be employed to minimize the high surface doses. Our proposed system is amendable to self-shielding and enhance research in regular laboratory setting.Keywords
This publication has 37 references indexed in Scilit:
- Experimental Platform for Ultra-high Dose Rate FLASH Irradiation of Small Animals Using a Clinical Linear AcceleratorInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2017
- Technical Note: Improvements in geant4 energy‐loss model and the effect on low‐energy electron transport in liquid waterMedical Physics, 2015
- Validation of the Geant4 simulation of bremsstrahlung from thick targets below 3 MeVNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2015
- Radiation dose to patients from X-ray radiographic examinations using computed radiography imaging systemJournal of Medical Physics, 2015
- Ultrahigh dose-rate FLASH irradiation increases the differential response between normal and tumor tissue in miceScience Translational Medicine, 2014
- A comprehensive system for dosimetric commissioning and Monte Carlo validation for the small animal radiation research platformPhysics in Medicine & Biology, 2009
- Benchmarking of Monte Carlo simulation of bremsstrahlung from thick targets at radiotherapy energiesMedical Physics, 2008
- Tumor therapy with heavy charged particlesProgress in Particle and Nuclear Physics, 2000
- The Constant Low Oxygen Concentration in All the Target Cells for Mouse Tail RadionecrosisRadiation Research, 1982
- Effects of Dose-rate on the Radiation Response of Rat SkinInternational Journal of Radiation Biology and Related Studies in Physics, Chemistry and Medicine, 1974