Cadaveric analysis of an accelerometer-based portable navigation device for distal femoral cutting block alignment in total knee arthroplasty
Authors:
Open Access
- 8 June 2012
- journal article
- research article
- Published by Taylor & Francis in Computer Aided Surgery
- Vol. 17 (4) , 205-210
Abstract
Femoral intramedullary guides have been shown to be insufficiently accurate in creating a distal femoral resection perpendicular to the mechanical axis in total knee arthroplasty (TKA), as they make assumptions regarding the difference between the patient's femoral mechanical and anatomical angles. The aim of this cadaveric study was to validate the accuracy of a portable accelerometer-based navigation device for alignment of the distal femoral cutting block in TKA. Twenty-nine trials were performed on five cadaveric specimens (hip-to-ankle), in which the distal femoral cutting block was placed using the KneeAlign 2™ navigation device. For each specimen, a preoperative “target” was assigned for varus/valgus and flexion/extension alignment of the cutting block. The actual alignment of each cutting block was then measured using the ORTHOsoft Computer Assisted Surgery (CAS) system. The mean absolute difference between the preoperative target and the alignment of the cutting block was 0.83 ± 0.60° for varus/valgus, and 0.83 ± 0.83° for flexion/extension. The KneeAlign 2™ navigation device can set and align the distal femoral resection guide with the same accuracy as a large-console CAS system, thus demonstrating that portable accelerometer-based navigation can be used reliably in total knee arthroplasty.Keywords
This publication has 18 references indexed in Scilit:
- Radiographic Analysis of a Hand-Held Surgical Navigation System for Tibial Resection in Total Knee ArthroplastyThe Journal of Arthroplasty, 2011
- Meta-Analysis of Alignment Outcomes in Computer-Assisted Total Knee Arthroplasty SurgeryThe Journal of Arthroplasty, 2007
- Comparison of Limb and Component Alignment Using Computer-Assisted Navigation Versus Image Intensifier–Guided Conventional Total Knee ArthroplastyThe Journal of Arthroplasty, 2007
- Mechanical Axis Cannot be Restored in Total Knee Arthroplasty With a Fixed Valgus Resection AngleThe Journal of Arthroplasty, 2007
- Alignments and Clinical Results in Conventional and Navigated Total Knee ArthroplastyClinical Orthopaedics and Related Research, 2007
- Current Etiologies and Modes of Failure in Total Knee Arthroplasty RevisionClinical Orthopaedics and Related Research, 2006
- Computer Assisted Navigation in Total Knee ArthroplastyThe Journal of Arthroplasty, 2005
- The Chetranjan Ranawat Award: Tibial Component Failure Mechanisms in Total Knee ArthroplastyClinical Orthopaedics and Related Research, 2004
- Why Are Total Knee Arthroplasties Failing Today?Clinical Orthopaedics and Related Research, 2002
- Extramedullary or intramedullary tibial alignment guides: a randomised, prospective trial of radiological alignmentThe Journal of Bone and Joint Surgery. British volume, 2002