Abstract
This work presents a fully disposable microchamber for gas generation of a sample solution. The microchamber consists of a cylindrical well-reactor and a paper-based microfluidic lid (μFluidic lid), which also serves as the reagent loading and dispensing unit. The base of the reactor consists of a hydrophobic membrane covering an in-house graphene electrochemical gas sensor. Fabrication of the gas sensor and the three-layer μFluidic lid is described. The μFluidic lid is designed to provide a steady addition of the acid reagent into the sample solution instead of liquid drops from a disposable syringe. There are three steps in the procedure: (i) acidification of the sample in the reactor to generate SO2 gas by the slow dispensing of the acid reagent from the μFluidic lid, (ii) diffusion of the liberated SO2 gas through the hydrophobic membrane at the base of the reactor, and (iii) in situ detection of SO2 by cathodic reduction at the graphene electrode. The device was demonstrated for quantitation of the sulfite preservative in wine without heating or stirring. The selectivity of the analysis is ensured by the combination of the gas-diffusion membrane and the selectivity of the electrochemical sensor. The linear working range is 2-60 mg L-1 SO2, with a limit of detection (3SD of intercept/slope) of 1.5 mg L-1 SO2. This in situ method has the shortest analysis time (8 min per sample) among all voltammetric methods that detect SO2(g) via membrane gas diffusion.Keywords
Funding Information
- National Research Council of Thailand (IRN/502/2563, NRCT5-RSA63015-01)
- Thailand Research Fund (PHD/0140/2558)
- Faculty of Science, Mahidol University
This publication has 31 references indexed in Scilit:
- In-line electrochemical reagent generation coupled to a flow injection biamperometric system for the determination of sulfite in beverage samplesFood Chemistry, 2016
- Free sulphite determination in wine using screen-printed carbon electrodes with prior gas-diffusion microextractionElectrochemistry Communications, 2016
- Carbon nanotubes for voltammetric determination of sulphite in some beveragesFood Chemistry, 2015
- Simple flow injection for determination of sulfite by amperometric detection using glassy carbon electrode modified with carbon nanotubes–PDDA–gold nanoparticlesTalanta, 2015
- Fast and reliable analyses of sulphite in fruit juices using a supramolecular amperometric detector encompassing in flow gas diffusion unitFood Chemistry, 2011
- Cathodic Reduction of Bisulfite and Sulfur Dioxide in Aqueous Solutions on Copper Electrodes: An Electrochemical ESR StudyThe Journal of Physical Chemistry B, 2005
- A novel electroreduction strategy for the determination of sulfiteThe Analyst, 2005
- Use of a copper electrode in alkaline medium as an amperometric sensor for sulphite in a flow-through configurationAnalytical and Bioanalytical Chemistry, 2002
- A review of sulphites in foods: Analytical methodology and reported findingsFood Additives & Contaminants, 1990
- The Reduction of Sulfurous Acid (Sulfur Dioxide) at the Dropping Mercury ElectrodeJournal of the American Chemical Society, 1941