A 14 MeV neutron spectrometer for the Joint European Torus deuterium-tritium experiments
Authors:
- 1 January 1999
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 70 (1) , 1134-1136
Abstract
A 14 MeV neutron spectrometer, based on the principle of the proton recoil telescope, has been installed as a deuterium–tritium (DT) plasma diagnostic at the Joint European Torus. The device comprises three identical telescope modules arranged one behind the other, to increase efficiency without worsening resolution. It views the plasma along a horizontal line of sight. The total efficiency is counts per (neutron per ), and the calculated response function has a full width at half maximum of 2.2% at an incident neutron energy of 14.1 MeV. The spectrometer was operational throughout the recent DT experiment No. 1, and selected neutron spectra have been analyzed in terms of the contributions from the various neutron production mechanisms.
Keywords
This publication has 4 references indexed in Scilit:
- Prediction/modelling of the neutron emission from hot-ion H-mode discharges in JETPlasma Physics and Controlled Fusion, 1997
- Simulations of deuterium-tritium experiments in TFTRNuclear Fusion, 1992
- The JET neutron time-of-flight spectrometerNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1991
- Fusion neutron energies and spectraPlasma Physics, 1973