Improvement of infrared imaging video bolometer for application to deuterium experiment on the large helical device
- 1 October 2018
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 89 (10) , 10E114
Abstract
An infrared imaging video bolometer was improved for application to a neutron environment in fusion plasma devices, i.e., the Large Helical Device (LHD). In order to calibrate the thermal characteristics of the activated foil absorber inside the plasma vacuum vessel, the remote-controlled in situ calibration system was improved with high-surface-flatness mirrors. Furthermore, the carbon coating method was improved by introducing a vacuum evaporation technique instead of the conventional spray technique to realize the coating on both sides of the absorber with reproducibility and uniformity. The optimal thickness of the coating was also determined. Owing to these coating improvements, the reproducibility of the effective emissivity on both sides especially was improved. Finally, the variation with the neutron irradiation of the thermal characteristics of the foil absorber was investigated. It was found that the effect was not significant for the total neutron emission of 3.6 × 1018 on LHD.Keywords
Funding Information
- National Institute for Fusion Science (NIFS16ULHH026)
- Japan Society for the Promotion of Science (JP17K14900)
This publication has 11 references indexed in Scilit:
- Experimental tests of an infrared video bolometer on Alcator C-ModReview of Scientific Instruments, 2018
- Tomographic reconstruction of two-dimensional radiated power distribution during impurity injection in KSTAR plasmas using an infrared imaging video bolometerCurrent Applied Physics, 2018
- In situ calibration of the foil detector for an infrared imaging video bolometer using a carbon evaporation techniqueReview of Scientific Instruments, 2016
- Development of impurity seeding and radiation enhancement in the helical divertor of LHDNuclear Fusion, 2015
- Inversion of infrared imaging bolometer based on one-dimensional and three-dimensional modeling in HL-2AReview of Scientific Instruments, 2014
- Improvement of Infrared Imaging Video Bolometer Systems in LHDPlasma and Fusion Research, 2014
- Foil Calibration for IR Imaging Bolometer by Laser IrradiationPlasma and Fusion Research, 2012
- Development of imaging bolometers for magnetic fusion reactors (invited)Review of Scientific Instruments, 2008
- Observation of divertor and core radiation in JT-60U by means of bolometric imagingJournal of Nuclear Materials, 2007
- Infrared imaging video bolometerReview of Scientific Instruments, 2000