Initial analytic quality assessment and method comparison of an immunoassay for adrenocorticotropic hormone measurement in equine samples
- 12 January 2016
- journal article
- research article
- Published by Wiley in Veterinary Clinical Pathology
- Vol. 45 (1) , 154-163
Abstract
Background: Equine pituitary pars intermedia dysfunction (PPID) may be diagnosed by measuring baseline plasma adrenocorticotrophic hormone (ACTH). The Immulite 1000 analyzer uses an automated chemiluminescence enzyme assay, previously validated for measuring equine ACTH. Recently, an automated bench‐top immunoassay analyzer (AIA‐360), designed for analytes in people, became available for veterinary use.Objectives: Objectives were to evaluate analytic performance of the AIA immunoassay for measuring equine ACTH, and compare the results with those obtained by the Immulite.Methods: Adrenocorticotrophic hormone was measured in plasma samples from 52 clinical cases. For the AIA, within‐ and between‐run coefficients of variation (CV) were assessed, linearity and recovery studies performed, and observed total error (TEobs) calculated. Correlation and agreement between the 2 analyzers were also evaluated.Results: Within‐run and between‐run CV of the AIA ranged from 2.3% to 4% and 3.5% to 8%, respectively. ACTH recoveries ranged from 89.5% to 115.9%. TEobs at 26.5 pg/mL ACTH was 4.1 pg/mL. The ACTH results (median: 25.9 pg/mL; range: 4.3–276.7 pg/mL) with AIA were significantly lower (P < .0001) than with the Immulite (median: 29.9 pg/mL; range: 10.3–639.0 pg/mL). Correlation between the 2 analyzers was r = 0.882 (P < .0001), with a significant bias for the AIA of −16 pg/mL. The 2 methods were not identical within inherent imprecision.Conclusion: The AIA is precise for measuring ACTH in horses. Although correlation between the instruments is good, the values obtained by the immunoassays cannot be used interchangeably and should be interpreted using reference intervals established for each analyzer to avoid false negatives. Diagnostic sensitivity and specificity of the AIA‐360 should be evaluated before clinical use.Keywords
Funding Information
- TOSOH Bioscience Ltd
This publication has 31 references indexed in Scilit:
- Comparison of Measurements of 12 Analytes in Equine Blood Samples Using the In-Practice Falcor 350 and the Reference KoneLab 30i AnalysersISRN Veterinary Science, 2012
- ASVCP quality assurance guidelines: control of general analytical factors in veterinary laboratoriesVeterinary Clinical Pathology, 2010
- Comparison of the VIDAS and IMMULITE‐2000 methods for cortisol measurement in canine serumVeterinary Clinical Pathology, 2009
- Correlation of Pituitary Histomorphometry with Adrenocorticotrophic Hormone Response to Domperidone Administration in the Diagnosis of Equine Pituitary Pars Intermedia DysfunctionVeterinary Pathology, 2008
- Preanalytical Stability of Adrenocorticotropic Hormone Depends on Time to Centrifugation Rather than TemperatureClinical Chemistry, 2007
- C-Reactive Protein Measurement in Canine SalivaJournal of Veterinary Diagnostic Investigation, 2005
- Nitration and Increased α‐Synuclein Expression Associated With Dopaminergic Neurodegeneration In Equine Pituitary Pars Intermedia DysfunctionJournal of Neuroendocrinology, 2005
- Validation of a Chemiluminescent Enzyme Immunometric Assay for Plasma Adrenocorticotropic Hormone in the DogVeterinary Clinical Pathology, 2003
- Treatment with Pergolide or Cyproheptadine of Pituitary Pars Intermedia Dysfunction (Equine Cushing's Disease)Journal of Veterinary Internal Medicine, 2002
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986