Abstract
Background: In recent years, there has been renewed interest in the “old” average of normals concept, now generally referred to as moving average quality control (MA QC) or patient-based real-time quality control (PBRTQC). However, there are some controversies regarding PBRTQC which this review aims to address while also indicating the current status of PBRTQC. Content: This review gives the background of certain newly described optimization and validation methods. It also indicates how QC plans incorporating PBRTQC can be designed for greater effectiveness and/or (cost) efficiency. Furthermore, it discusses controversies regarding the complexity of obtaining PBRTQC settings, the replacement of iQC, and software functionality requirements. Finally, it presents evidence of the added value and practicability of PBRTQC. Outlook: Recent developments in, and availability of, simulation methods to optimize and validate laboratory-specific PBRTQC procedures have enabled medical laboratories to implement PBRTQC in their daily practice. Furthermore, these methods have made it possible to demonstrate the practicability and added value of PBRTQC by means of two prospective “clinical” studies and other investigations. Although internal QC will remain an essential part of any QC plan, applying PBRTQC can now significantly improve its performance and (cost) efficiency.Keywords
This publication has 36 references indexed in Scilit:
- Patient pools and the use of “patient means” are valuable tools in quality control illustrated by a bone-specific alkaline phosphatase assayClinical Chemistry and Laboratory Medicine (CCLM), 2016
- Monitoring laboratory data across manufacturers and laboratories—A prerequisite to make “Big Data” workClinica Chimica Acta, 2015
- Changing the paradigm of laboratory quality control through implementation of real-time test results monitoring: For patients by patientsClinical Biochemistry, 2015
- The Empower project – a new way of assessing and monitoring test comparability and stabilityClinical Chemistry and Laboratory Medicine (CCLM), 2015
- Commutability Limitations Influence Quality Control Results with Different Reagent LotsClinical Chemistry, 2011
- Practical application of Sigma Metrics QC procedures in clinical chemistryClinical Chemistry and Laboratory Medicine (CCLM), 2011
- Performance Evaluation and Planning for Patient-Based Quality Control ProceduresAmerican Journal of Clinical Pathology, 2000
- Assessment of “Average of Normals” Quality Control Procedures and Guidelines for ImplementationAmerican Journal of Clinical Pathology, 1984
- A Study of Various Estimators for the Derivation of Quality Control Procedures from Patient Erythrocyte IndicesAmerican Journal of Clinical Pathology, 1974
- The “Average of Normals” Method of Quality ControlAmerican Journal of Clinical Pathology, 1965