Increasing sensitivity and decreasing spot size using an inexpensive, removable hydrophobic coating for matrix‐assisted laser desorption/ionisation plates
- 7 October 2003
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 17 (21) , 2439-2449
Abstract
Spot size reduction and increased detection sensitivity in matrix‐assisted laser desorption/ionisation (MALDI) of small molecules are accomplished by using an inexpensive and removable hydrophobic coating for MALDI targets, based on 3M Scotch Gard™ surface treatment. Several variations in sample preparation were explored, such as surface coating technique, identity of the matrix, solvent composition, and the type of metal support plate used. These were investigated on both uncoated and coated surfaces and their impact on spot size, crystal coverage, and sensitivity is presented here. Additionally, crystallisation behaviour obtained on coated plates is compared with that on uncoated plates using scanning electron microscope analysis. To demonstrate the potential of the new coating technique, erythromycin A and valinomycin are studied to determine the increase in detection sensitivity of coated plates in comparison to uncoated plates, and to reveal the suitability of the plates for application in combined high‐performance liquid chromatography/MALDI (HPLC/MALDI), where widely varying solvent compositions and droplet volumes are observed. It is shown that enhancements in detection sensitivities correlate very well with the achieved spot size reduction. The versatility of the coated plates is also exhibited by the ease of removing the surface layer, after which the plates can be rigorously cleaned without worry about damaging the hydrophobic surface, followed by a quick reapplication of new hydrophobic coating material. This makes the non‐polar coating superior to more expensive commercial hydrophobic‐coated targets, which are much more delicate to clean. Furthermore, cleaning and reapplication eliminate potential carry‐over effects and the easy application procedure also makes the fabrication of inexpensive, disposable MALDI targets readily possible. Copyright © 2003 John Wiley & Sons, Ltd.Keywords
This publication has 27 references indexed in Scilit:
- An atmospheric pressure matrix‐assisted laser desorption/ionization ion trap with enhanced sensitivityRapid Communications in Mass Spectrometry, 2003
- A new method to improve sensitivity and resolution in matrix‐assisted laser desorption/ionization time of flight mass spectrometryProteomics, 2003
- Derivatization of small biomolecules for optimized matrix‐assisted laser desorption/ionization mass spectrometryJournal of Mass Spectrometry, 2002
- One‐way hydrophobic surface foil for UV matrix‐assisted laser desorption/ionization mass spectrometry of peptidesRapid Communications in Mass Spectrometry, 2002
- Hyphenation of capillary high‐performance liquid chromatography with matrix‐assisted laser desorption/ionization time‐of‐flight mass spectrometry for nano‐scale screening of single‐bead combinatorial librariesRapid Communications in Mass Spectrometry, 2002
- MALDI-TOF-MS Analysis of Droplets Prepared in an Electrodynamic Balance: “Wall-less” Sample PreparationAnalytical Chemistry, 2001
- Ready‐made matrix‐assisted laser desorption/ionization target plates coated with thin matrix layer for automated sample deposition in high‐density array formatRapid Communications in Mass Spectrometry, 2001
- Accurate mass measurement of synthetic analogues of prazosine by matrix‐assisted laser desorption/ionisation time‐of‐flight mass spectrometryRapid Communications in Mass Spectrometry, 2001
- Erythromycin: New chemistry on an old compound1994
- Matrix-assisted ultraviolet laser desorption of non-volatile compoundsInternational Journal of Mass Spectrometry and Ion Processes, 1987