Abstract
This study presents a laboratory-scale investigation of a nano-enhanced smart bedsheet designed for continuous, non-invasive patient monitoring. The work focuses on evaluating the electrical behavior, sensitivity, response time, and stability of textile-embedded sensors integrated with silver nanoparticle (AgNP) antimicrobial coatings and a hydrophilic polyvinyl alcohol (PVA) finish. Pressure-related motion, moisture presence, and colorimetric fluid detection were monitored using a MEMS accelerometer, a polyaniline-based moisture sensor, and silicon photodiodes with RGB filters, respectively. Experimental calibration was conducted using controlled mechanical loads and liquid volumes, and performance metrics including sensitivity, response time, repeatability, and drift were quantified. Results demonstrate that AgNP-coated textiles exhibit improved signal stability and reduced baseline noise compared to uncoated fabric, while PVA finishing enhances liquid absorption uniformity. The system provides near real-time alerts through a microcontroller-GSM interface under laboratory conditions. The proposed bedsheet is intended as a pre-clinical prototype and has not yet been validated against medical device standards.

This publication has 12 references indexed in Scilit: