Is Cartilage Interposition Required for Ossiculoplasty With Titanium Prostheses?
- 1 March 2014
- journal article
- prosthetic devices
- Published by Wolters Kluwer Health in Otology & Neurotology
- Vol. 35 (3) , 482-488
Abstract
Objective To demonstrate that cartilage interposition between the head of a titanium prosthesis and the tympanic membrane is not necessary to prevent extrusion. Study Design Retrospective review. Setting District general hospital. Patients A retrospective review of 73 ears in patients aged between 5.9 and 64.5 years, which had an ossiculoplasty performed using a titanium prosthesis between the stapes footplate or stapes head and the tympanic membrane. Interventions Use of Spiggle and Theis two part, shortenable titanium PORP and TORP and Grace Alto titanium TORP and PORP. Main Outcome Measures Number of extruded prostheses. Audiometric results. Results Five extrusions occurred in 4 patients, follow-up between 6 and 98 months. Extrusions were related to retraction of the tympanic membrane around the prosthesis. Audiometric results compared favorably with other ossiculoplasty techniques. Conclusion The titanium head of the Spiggle and Theis and Grace Alto titanium prostheses interact well with the undersurface of the tympanic membrane in patients with favorable middle ear conditions. In our experience, cartilage interposition is not mandatory.Keywords
This publication has 4 references indexed in Scilit:
- A Micro-Computed Tomographic StudyOtology & Neurotology, 2011
- A Modified Prosthesis Head for Middle Ear Titanium Implants-Experimental and First Clinical ResultsOtology & Neurotology, 2010
- A New L-Shaped Titanium Prosthesis for Total Reconstruction of the Ossicular ChainOtology & Neurotology, 2004
- Gehörknöchelchenersatz mit Titan-ProthesenLaryngo-Rhino-Otologie, 1996