Quality assurance of volumetric modulated arc therapy using Elekta Synergy
Authors:
- Akihiro HagaAHAkihiro Haga
- Keiichi NakagawaKNKeiichi Nakagawa
- Kenshiro ShiraishiKSKenshiro Shiraishi
- Saori ItohSISaori Itoh
- Atsuro TeraharaATAtsuro Terahara
- Hideomi YamashitaHYHideomi Yamashita
- Kuni OhtomoKOKuni Ohtomo
- Shigeki SaegusaSSShigeki Saegusa
- Toshikazu ImaeTIToshikazu Imae
- Kiyoshi YodaKYKiyoshi Yoda
- Roberto PellegriniRPRoberto Pellegrini
- 1 January 2009
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Oncologica
- Vol. 48 (8) , 1193-1197
Abstract
Purpose. Recently, Elekta has supplied volumetric modulated arc therapy (VMAT) in which multi-leaf collimator (MLC) shape, jaw position, collimator angle, and gantry speed vary continuously during gantry rotation. A quality assurance procedure for VMAT delivery is described. Methods and materials. A single-arc VMAT plan with 73 control points (CPs) and 5-degree gantry angle spacing for a prostate cancer patient has been created by ERGO + + treatment planning system (TPS), where MLC shapes are given by anatomic relationship between a target and organs at risk and the monitor unit for each CP is optimized based on given dose prescriptions. Actual leaf and jaw positions, gantry angles and dose rates during prostate VMAT delivery were recorded in every 0.25 seconds, and the errors between planned and actual values were evaluated. The dose re-calculation using these recorded data has been performed and compared with the original TPS plan using the gamma index. Results. Typical peak errors of gantry angles, leaf positions, and jaw positions were 3 degrees, 0.6 mm, and 1 mm, respectively. The dose distribution obtained by the TPS plan and the recalculated one agreed well under 2%-2 mm gamma index criteria. Conclusions. Quality assurance for prostate VMAT delivery has been performed with a satisfied result.Keywords
This publication has 9 references indexed in Scilit:
- Commissioning of Volumetric Modulated Arc Therapy (VMAT)International Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Dosimetric verification of RapidArc treatment deliveryActa Oncologica, 2009
- Treatment of lung cancer using volumetric modulated arc therapy and image guidance: A case studyActa Oncologica, 2008
- Volumetric modulated arc therapy: IMRT in a single gantry arcMedical Physics, 2007
- An arc‐sequencing algorithm for intensity modulated arc therapyMedical Physics, 2007
- Inverse planning for intensity-modulated arc therapy using direct aperture optimizationPhysics in Medicine & Biology, 2003
- Clinical implementation of intensity-modulated arc therapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2002
- Intensity-modulated arc therapy simplifiedInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2002
- Intensity-modulated arc therapy with dynamic multileaf collimation: an alternative to tomotherapyPhysics in Medicine & Biology, 1995