Commutability of a Whole-Blood External Quality Assessment Material for Point-of-Care C-Reactive Protein, Glucose, and Hemoglobin Testing
Authors:
- 1 June 2019
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 65 (6) , 791-797
Abstract
BACKGROUND: The optimal situation in external quality assessment (EQA) is to use commutable materials. No previous study has examined the commutability of a whole-blood material for point-of-care (POC) testing. The aim of this study was to determine the commutability of the Norwegian Quality Improvement of Laboratory Examinations (Noklus) organization's “in-house” whole-blood EQA material for C-reactive protein (CRP), glucose, and hemoglobin for frequently used POC instruments in Norway and to determine the possibility of using a common target value for each analyte.METHODS: The study was performed according to the Clinical and Laboratory Standards Institute guidelines. The EQA material was pooled stabilized EDTA venous whole-blood containing different concentrations of the analytes. The EQA material and native routine patient samples were analyzed using 17 POC and 3 hospital instruments. The commutability was assessed using Deming regression analysis with 95% prediction intervals for each instrument comparison.RESULTS: The EQA material was commutable for all CRP and hemoglobin POC instruments, whereas for glucose the material was commutable for all POC instruments at the lowest concentration analyzed [126.0 mg/dL (7.0 mmol/L)] and for 3 POC instruments at all of the concentrations analyzed.CONCLUSIONS: Noklus EQA participants using CRP and hemoglobin POC instruments now receive results that are compared with a reference target value, whereas the results for participants using glucose POC instruments are still compared with method-specific target values. Systematic deviations from a reference target value for the commutable glucose POC instruments can be calculated, and this additional information can now be offered to these participants and to the manufacturers.Keywords
This publication has 14 references indexed in Scilit:
- Essential aspects of external quality assurance for point-of-care testingBiochemia Medica, 2017
- The Importance of Reagent Lot Registration in External Quality Assurance/Proficiency Testing SchemesClinical Chemistry, 2016
- Harmonization: the Sample, the Measurement, and the ReportAnnals of Laboratory Medicine, 2014
- How to achieve harmonisation of laboratory testing —The complete pictureClinica Chimica Acta, 2014
- Application of the Reference Method Isotope Dilution Gas Chromatography Mass Spectrometry (ID/GC/MS) to Establish Metrological Traceability for Calibration and Control of Blood Glucose Test SystemsJournal of Diabetes Science and Technology, 2014
- Proficiency Testing/External Quality Assessment: Current Challenges and Future DirectionsClinical Chemistry, 2011
- Analytical performance evaluation of the Cobas 6000 analyzer – special emphasis on trueness verificationClinical Chemistry and Laboratory Medicine (CCLM), 2008
- Between-Lot Variation in External Quality Assessment of Glucose: Clinical Importance and Effect on Participant Performance EvaluationClinical Chemistry, 2005
- Evaluation of Commutability of Control MaterialsClinical Chemistry and Laboratory Medicine (CCLM), 2002
- External quality assessment in primary health care by using fresh whole bloodClinical Chemistry, 1994