Prospective Gating With 320-MDCT Angiography: Effect of Volume Scan Length on Radiation Dose

Authors:

Abstract
OBJECTIVE. The purpose of this study was to evaluate the relation between radiation dose reduction and volume scan length for prospectively ECG-gated 320-MDCT angiography in the diagnosis of coronary artery disease.MATERIALS AND METHODS. MDCT with prospective ECG gating was performed at one of the three volume scan lengths depending on heart length. Of 175 patients, 95 (55%; body mass index, 29 ± 5.9; mean heart rate, 59 ± 7 beats/min) underwent scanning at 160 mm; 46 (26%; body mass index, 30 ± 4.1; mean heart rate, 56 ± 5.74 beats/min) at 140 mm; and 34 (19%; body mass index, 30 ± 3.71; mean heart rate, 58 ± 3.96 beats/min) at 120 mm.RESULTS. The median radiation doses were 6.5 mSv (95% CI, 6.03–7.2 mSv) for the 95 patients who underwent scanning at a volume scan length of 160 mm, 4.33 mSv (95% CI, 4.06–6.62 mSv) for the 46 patients who underwent scanning at 140 mm, and 3.47 mSv (95% CI, 3.15–3.62 mSv) for the 34 patients who underwent scanning at 120 mm. The reduction in scan length from 160 to 140 mm ...