Neutron field characterisation of the OB26 CRNA irradiator in view of its use for calibration purposes
Authors:
- 16 June 2010
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 141 (2) , 114-126
Abstract
The main goal of the present work is to characterise the neutron field of an OB26 irradiation system acquired by the Nuclear Research Center of Algiers for radiation protection purposes. Extensive Monte-Carlo (MC) calculations and measurements using BF3- and 3He-based neutron area dosemeters were performed to estimate the contribution, on the energy neutron spectrum, of each component present in the bunker facility of the Algerian Secondary Standard Dosimetry Laboratory (SSDL) where the irradiator has been installed. For this purpose, new irradiation configurations based on the 241Am–Be source placed in the OB 26/2 biological shielding inside its environment have been investigated by MC simulations, and comparison with the ISO spectrum has been performed. During MC simulations, sensitivity analysis has been considered to estimate the effect of several physical parameters on the neutron fluence and dose equivalent rates. In addition, the contribution of the gamma dose equivalent rates to the total neutron dose equivalent rates was estimated for both selected source–detector distances (SDDs) 150 and 200 cm. Finally, a theoretical approach has been adopted, using MCNP5 fluence rates, to estimate the readings of the instruments taking into account their response functions. A low mean difference (12 %) between measured and predicted dose equivalent rates for two selected SDDs has been observed. Overall, the obtained MCNP5 results regarding the actual SSDL irradiation facility are particularly encouraging, but need to be supported by further experimental data.Keywords
This publication has 10 references indexed in Scilit:
- Characterization of a sealed Americium–Beryllium (AmBe) source by inductively coupled plasma mass spectrometryJournal of Radioanalytical and Nuclear Chemistry, 2009
- Experimental comparison of 241Am-Be Neutron fluence energy distributionsRadiation Protection Dosimetry, 2007
- Investigation of the neutron energy distribution of the IRSN 241Am-Be( ,n) sourceRadiation Protection Dosimetry, 2006
- Characteristics of the neutron field of the facility at DIN-UPMRadiation Protection Dosimetry, 2004
- Monte Carlo simulation of the IRSN CANEL/T400 realistic mixed neutron-photon radiation fieldRadiation Protection Dosimetry, 2004
- Measurement and Monte Carlo modelling of the JRC 241Am-Li( ,n) source spectrumRadiation Protection Dosimetry, 2004
- Characterisation of the IRSN graphite moderated Americium-Beryllium neutron fieldRadiation Protection Dosimetry, 2004
- Neutron spectrometry in mixed fields: multisphere spectrometersRadiation Protection Dosimetry, 2003
- Measurement of the 241Am–Li radionuclide neutron source spectrumNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- The Neutron Fluence and H*(10) Response of the New LB 6411 REM CounterRadiation Protection Dosimetry, 1997