Relative dosimetry in a 1.5 T magnetic field: an MR-linac compatible prototype scanning water phantom
Authors:
- J SjöholmJSJ Sjöholm
- J G M KokJKJ G M Kok
- J J W LagendijkJLJ J W Lagendijk
- B W RaaymakersBRB W Raaymakers
- 3 July 2014
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 59 (15) , 4099-4109
Abstract
The MR-linac is a hybrid MRI radiotherapy system allowing dose delivery in a 1.5 T magnetic field. This paper presents the design and performance of a prototype MR-linac compatible scanning water phantom. Since a scanning water phantom requires dose detectors, the performance air-filled ionization chambers in the magnetic field was characterized. We have found that the linearity and reproducibility of an ionization chamber are unaffected by the magnetic field. Also, moving the ionization chambers in a magnetic field during irradiation does not affect the dose response. When scanning in-plane profiles, the change in irradiation orientation can influence the ionization chamber dose response by up to 0.4%. However this effect can be eliminated by rotating the ionization chamber by 90° before measuring the in-plane profile. The performance of the total scanning water phantom was validated at a clinical setup in a 0 T magnetic field. There was no significant difference between the dose profiles measured with a standard clinical scanning water phantom and the profiles measured with the MR-linac compatible scanning water phantom. The performance of the MR-linac scanning water phantom in the MR-linac was validated using Gafchromic EBT2 film. There was no significant difference in dose profiles between the MR-linac scanning water phantom and the radiochromic film. These results indicate that automated scanning water phantom measurements using ionization chamber detectors are possible in the MR-linac.Keywords
This publication has 12 references indexed in Scilit:
- Electron contamination modeling and skin dose in 6 MV longitudinal field MRIgRT: Impact of the MRI and MRI fringe fieldMedical Physics, 2012
- Integrating a 1.5 T MRI scanner with a 6 MV accelerator: proof of conceptPhysics in Medicine & Biology, 2009
- First MR images obtained during megavoltage photon irradiation from a prototype integrated linac‐MR systemMedical Physics, 2009
- Dosimetry for the MRI accelerator: the impact of a magnetic field on the response of a Farmer NE2571 ionization chamberPhysics in Medicine & Biology, 2009
- Accelerator beam data commissioning equipment and procedures: Report of the TG‐106 of the Therapy Physics Committee of the AAPMMedical Physics, 2008
- MRI/linac integrationRadiotherapy and Oncology, 2008
- Experimental verification of magnetic field dose effects for the MRI-acceleratorPhysics in Medicine & Biology, 2007
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: impact of the surface orientation on the entrance and exit dose due to the transverse magnetic fieldPhysics in Medicine & Biology, 2007
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue–air interfaces in a lateral magnetic field due to returning electronsPhysics in Medicine & Biology, 2005
- Acceptance testing of an automated scanning water phantomMedical Physics, 1990