Woven EndoBridge device for the treatment of ruptured intracranial aneurysms: A systematic review of clinical and angiographic results
- 23 June 2021
- journal article
- review article
- Published by SAGE Publications in Interventional Neuroradiology
- Vol. 28 (2) , 240-249
Abstract
Background and objective: The clinical outcome and angiographic outcome data of Woven EndoBridge (WEB) device for the treatment of ruptured intracranial aneurysms (IAs) are limited. We conducted a meta-analysis of the latest literature on the WEB device in the treatment of ruptured IAs. Methods: A comprehensive literature search of 4 databases (PubMed, Web of Science, Cochrane library, and Embase) was conducted for studies published from January 1, 2010 to December 31, 2020. Two reviewers independently extracted variables (aneurysm and patient characteristics) using a prespecified data-collection sheet. Outcomes studied included initial and latest follow-up angiographic outcomes, technical success rate, perioperative mortality, retreated rate, perioperative re-bleeding, complication, intraoperative rupture, favorable neurologic outcome at discharge. We used random-effects model to pool the data. Results: We finally presented the results of 7 articles including 276 patients with 283 aneurysms. Initial complete and adequate occlusion rate were 38% (95% CI, 25%–50%) and 98% (95% CI, 95%–100%), respectively. Latest follow-up complete and adequate occlusion rate were 61% (95% CI, 46%–75%) and 91% (95% CI, 84%–98%), respectively.Technical success rate was 99% (95% CI, 98%–100%). Perioperative mortality rates and perioperative re-bleeding rate were 9% (95% CI, 3%–15%) and 1% (95% CI, 0%–2%), respectively. Retreated rate was 6% (95% CI, 3%–10%). Overall and WEB treatment-related thromboembolic complication was 10% (95% CI, 6%–13%) and 7% (95% CI, 2%–12%), respectively. Intraoperative rupture rate was 3% (95% CI, 0%–6%). Conclusion: Endovascular treatment of ruptured IAs with the WEB device has a good safety profile and an acceptable aneurysm occlusion rate.Keywords
This publication has 35 references indexed in Scilit:
- Endovascular Treatment of Intracranial Aneurysms With Flow DivertersStroke, 2013
- Stent-Assisted Coiling of Wide-Necked Aneurysms in the Setting of Acute Subarachnoid HemorrhageNeurosurgery, 2012
- Stent assisted coiling of the ruptured wide necked intracranial aneurysmJournal of NeuroInterventional Surgery, 2011
- The Woven EndoBridge Cerebral Aneurysm Embolization Device (WEB II): initial clinical experienceNeuroradiology, 2011
- The Woven EndoBridge: A New Aneurysm Occlusion DeviceAmerican Journal of Neuroradiology, 2011
- Detachable coil embolisation of ruptured intracranial aneurysms: A single center study, a decade experienceClinical Neurology and Neurosurgery, 2009
- Incidence of subarachnoid haemorrhage: a systematic review with emphasis on region, age, gender and time trendsJournal of Neurology, Neurosurgery & Psychiatry, 2007
- A New Endoluminal, Flow-Disrupting Device for Treatment of Saccular AneurysmsStroke, 2007
- A Multinational Comparison of Subarachnoid Hemorrhage Epidemiology in the WHO MONICA Stroke StudyStroke, 2000
- Meta-analysis in clinical trialsControlled Clinical Trials, 1986