An empirical method for the determination of wall perturbation factors for parallel-plate chambers in high-energy electron beams
- 27 September 2006
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 51 (20) , 5167-5181
Abstract
The calibration of ion chambers in high-energy electron beams in terms of absorbed dose to water at the National Physical Laboratory requires knowledge of the ratio of perturbation factors in graphite and water phantoms. During a review of data required for the NPL calibration procedure an empirical model was developed to calculate the perturbation due to the rear wall, pwall, of a well-guarded ion chamber in a high-energy electron beam. The overall uncertainty in this method is estimated to be 0.4%, which is the lowest value reported to date. The model reproduces measured data at the 0.1% level or better and indicates that the NACP ion chamber has a nonzero perturbation factor in electron beams due to backscatter from the rear wall. The effect is small (4 cm, E0>10 MeV) but becomes large at low energies-up to 1.4% at E0=4 MeV (R50=1.2 cm). The model indicates that there is a nonzero correction for the NACP chamber in both a graphite and water phantom and that material adjacent to the air cavity has a significant effect on the measured ionization. These values are consistent with previous measurements and recent Monte Carlo calculations. The model could be used in the design of ion chambers and in the estimation of corrections for non-homogeneous systems, especially in the absence of accurate Monte Carlo simulations.Keywords
This publication has 17 references indexed in Scilit:
- Wall correction factors, Pwall, for parallel-plate ionization chambersMedical Physics, 2006
- Characterization of the phantom material Virtual Water™ in high‐energy photon and electron beamsMedical Physics, 2006
- The IPEM code of practice for electron dosimetry for radiotherapy beams of initial energy from 4 to 25 MeV based on an absorbed dose to water calibrationPhysics in Medicine & Biology, 2003
- Determination of absorbed dose calibration factors for therapy level electron beam ionization chambersPhysics in Medicine & Biology, 2001
- Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 versionMedical Physics, 2000
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- The calibration of therapy level electron beam ionization chambers in terms of absorbed dose to waterPhysics in Medicine & Biology, 1998
- The IPEMB code of practice for electron dosimetry for radiotherapy beams of initial energy from 2 to 50 MeV based on an air kerma calibrationPhysics in Medicine & Biology, 1996
- Implication of electron backscattering for electron dosimetryPhysics in Medicine & Biology, 1991
- Electron backscatter corrections for parallel‐plate chambersMedical Physics, 1988