A Simple Low-Cost Wearable Sensor for Long-Term Ambulatory Monitoring of Knee Joint Kinematics
- 19 November 2020
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 67 (12) , 3483-3490
Abstract
Objective: Accurate monitoring of joint kinematics in individuals with neuromuscular and musculoskeletal disorders within ambulatory settings could provide important information about changes in disease status and the effectiveness of rehabilitation programs and/or pharmacological treatments. This paper introduces a reliable, power efficient, and low-cost wearable system designed for the long-term monitoring of joint kinematics in ambulatory settings. Methods: Seventeen healthy subjects wore a retractable string sensor, fixed to two anchor points on the opposing segments of the knee joint, while walking at three different self-selected speeds. Joint angles were estimated from calibrated sensor values and their derivatives in a leave-one-subject-out cross validation manner using a random forest algorithm. Results: The proposed system estimated knee flexion/extension angles with a root mean square error (RMSE) of 5.0° ± 1.0° across the study subjects upon removal of a single outlier subject. The outlier was likely a result of sensor miscalibration. Conclusion: The proposed wearable device can accurately estimate knee flexion/extension angles during locomotion at various walking speeds. Significance: We believe that our novel wearable technology has great potential to enable joint kinematic monitoring in ambulatory settings and thus provide clinicians with an opportunity to closely monitor joint recovery, develop optimal, personalized rehabilitation programs, and ultimately maximize therapeutic outcomes.Keywords
Funding Information
- National Institute of Biomedical Imaging and Bioengineering (R21EB025284)
This publication has 37 references indexed in Scilit:
- A real-time system for biomechanical analysis of human movement and muscle functionMedical & Biological Engineering & Computing, 2013
- Ambulatory Monitoring of Physical Activity Based on Knee Flexion/Extension Measured by Inductive Sensor TechnologyISRN Biomedical Engineering, 2013
- A review of wearable sensors and systems with application in rehabilitationJournal of NeuroEngineering and Rehabilitation, 2012
- The use of near-infrared spectroscopy in understanding skeletal muscle physiology: recent developmentsPhilosophical Transactions of the Royal Society A: Mathematical, Physical and Engineering Sciences, 2011
- A durable, low-cost electrogoniometer for dynamic measurement of joint trajectoriesMedical Engineering & Physics, 2011
- The association between knee joint biomechanics and neuromuscular control and moderate knee osteoarthritis radiographic and pain severityOsteoarthritis and Cartilage, 2011
- Gait and neuromuscular adaptations after using a feedback-based gait monitoring knee braceGait & Posture, 2010
- Inertial sensor-based knee flexion/extension angle estimationJournal of Biomechanics, 2009
- Indications for Rotator Cuff RepairClinical Orthopaedics and Related Research, 2007
- Effect of auditory rhythmic cuing on gait kinematic parameters of stroke patientsGait & Posture, 1997