Abstract
Background Determination of fractional flow reserve (FFR) has been proposed as a means to assess stent deployment. In this prospective, multicenter trial, we evaluate the use of FFR to optimize stenting by comparing it with standard intravascular ultrasound (IVUS) criteria. Methods and Results Eighty-four stable patients with isolated coronary lesions underwent coronary stent deployment starting at 10 atm and increased serially by 2 atm until the FFR was ≥0.94 or 16 atm was achieved. IVUS was then performed. FFR was measured with a coronary pressure wire with intracoronary adenosine to induce hyperemia. The diagnostic characteristics of an FFR 30 μg) of intracoronary adenosine during pressure measurements, suggesting that FFR might be overestimated in the other group. Conclusions A fractional flow reserve <0.96, measured after stent deployment, predicts a suboptimal result based on validated intravascular ultrasound criteria; however, an FFR ≥0.96 does not reliably predict an optimal stent result. Higher doses of intracoronary adenosine than previously used to measure FFR improve these results.